A new reagent carrier, Reflotron ALP, has been developed for the Reflotron system, allowing easy and rapid measurement (in less than 3 minutes) of alkaline phosphatase (ALP) activity in capillary blood, venous blood, heparinized plasma or serum. The evaluation of the analytical performance of the assay was carried out at eight clinical laboratories. The study of the imprecision using the measurements in human samples resulted in coefficients of variation ranging from 1.3% to 4.6% (within-run) and from 3.2% to 4.0% (day-to-day). The analytical specificity of the Reflotron ALP assay agrees well with ALP methods using a N-methyl-D-glucamine buffer solution. The calibration of the Reflotron ALP assay, however, is related to the reference intervals for ALP methods using a diethanolamine buffer solution. Method comparisons were performed with the ALP method on Hitachi instruments using diethanolamine buffer. Reflotron ALP measurements in blood and plasma in 157 randomly selected split samples showed excellent agreement (slope: 0.99; intercept: 0.7 U/l; median bias: 2.3%; median difference from the comparison method: -0.3%). Specimens from pregnant women and adolescents were excluded from this study. Differing values were obtained in a method comparison using 48 samples containing predominantly the ALP bone isoform (slope: 0.81; intercept: 31.5 U/l; median bias: 5.7%; median difference from the comparison method: -12.2%). Regression analysis of the results from 21 sera with prevailing placental ALP gave a slope of 1.51, and an intercept of -41.1 U/l (median bias: 8.6%; median difference from the comparison method: 35.6%). Reflotron ALP was compared with three different wet chemistry procedures using different buffer compounds: N-methyl-D-glucamine or diethanolamine or 2-amino-2-methyl-1-propanol. In samples containing predominantly ALP isoforms not of liver origin, the measurements with N-methyl-D-glucamine buffer gave the best fit with respect to Reflotron. In an interference study with 18 drugs, no effect on the test results could be detected. Total bilirubin up to 750 micromol/l and hemolysis up to 1.7 g/l free hemoglobin did not influence the test. Reflotron ALP proved to be an easy and rapid method with excellent precision. The accuracy related to an ALP method using diethanolamine buffer was good. The systematic differences for ALP in samples from pregnant women and adolescents have to be taken into account. The assay is well suited for differential diagnosis of hepatic diseases in decentralized testing.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1515/CCLM.2001.015 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!