R-wave detection by subcutaneous ECG. Possible use for analyzing R-R variability.

Ann Noninvasive Electrocardiol

Medical Clinic I, Charité Hospital, Schumannstr. 20-21, 10089 Berlin, Germany.

Published: January 2001

AI Article Synopsis

Article Abstract

Background: Atrial arrhythmia (AA) discrimination remains a technological challenge for implanted cardiac devices. We examined the feasibility of R-wave detection by a subcutaneous far field ECG (SFFECG) and analysis of these signals for R to R variability as an indicator of atrial arrhythmia (AA).

Methods: Surface ECG and SFFECG (from the pacemaker pocket) were recorded in sixteen patients (61.5 +/- 11.4 years) with AA. The SFFECG was recorded with a pacemaker sized four electrode array acutely placed in the pacemaker pocket during implantation. The signals were analyzed to obtain peak-to-peak R wave amplitude and R to R interval variability (indicative of AAs).

Results: In sixteen patients R waves were visually discernible in all recordings. The percentage over and under detection for automatic R wave recognition SFFECG was 3 and 9%, respectively. R to R variability analysis using the SFFECG produced results concordant to those using the surface ECG.

Conclusion: SFFECG might be a helpful adjunct in implantable device systems for detection of R waves and may be used for measurement of R to R variability.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7027661PMC
http://dx.doi.org/10.1111/j.1542-474x.2001.tb00081.xDOI Listing

Publication Analysis

Top Keywords

r-wave detection
8
detection subcutaneous
8
atrial arrhythmia
8
ecg sffecg
8
pacemaker pocket
8
sixteen patients
8
sffecg
6
variability
5
subcutaneous ecg
4
ecg analyzing
4

Similar Publications

Current guidelines recommend cardioverter-defibrillator (ICD) programming, including faster detection rates, longer detection durations, and strict discrimination for supraventricular tachycardia (SVT) to prevent unnecessary ICD treatment. This delayed-style ICD programming could lead to a rise in the possibility of VF undersensing. To avoid this risk, an innovative algorithm known as VF Therapy Assurance (VFTA; Abbott, Sylmar, CA) has been developed.

View Article and Find Full Text PDF
Article Synopsis
  • Continuous electrocardiographic (ECG) monitoring during surgery in infants and children is essential to detect potential myocardial pathologies, despite the rarity of these issues in pediatric patients.
  • A case is presented involving a 2-year-old child who showed ST segment depression and increased R wave amplitude during surgery, which led to the discovery of undiagnosed hypertrophic cardiomyopathy postoperatively.
  • The findings emphasize the importance of recognizing ECG changes and the need for thorough evaluation to prevent complications during and after surgery, even when patients appear asymptomatic.
View Article and Find Full Text PDF

The sympathetic nervous system (SNS) is essential for the body's immediate response to stress, initiating physiological changes that can be measured through sympathetic nerve activity (SNA). While microneurography (MNG) is the gold standard for direct SNA measurement, its invasive nature limits its practical use in clinical settings. This study investigates the use of multi-wavelength photoplethysmography (PPG) as a non-invasive alternative for SNA measurement.

View Article and Find Full Text PDF

We report 3 cases of irregular, narrow complex tachycardia misdiagnosed and treated for atrial fibrillation. The adenosine response, detection of recurring triple cycle length variation patterns, and pseudo-R-wave in lead V during tachycardia made us suspect typical atrioventricular nodal re-entrant tachycardia. The electrophysiology study confirmed atrioventricular nodal re-entrant tachycardia, and symptoms were resolved by slow pathway modification.

View Article and Find Full Text PDF

Introduction: The novel Confirm Rx™ implantable cardiac monitor (ICM) with SharpSense™ technology incorporates a new P-wave discriminator designed to improve AF detection. This study aimed to evaluate the diagnostic performance of the Confirm Rx™ ICM in detecting AF episodes of varying durations.

Methods: We conducted a multicenter retrospective analysis of consecutive patients implanted with a Confirm Rx™ ICM (v1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!