Spinning Drop Tensiometry Using a Square Section Sample Tube.

J Colloid Interface Sci

Department of Civil and Environmental Engineering, Massachusetts Institute of Technology, Cambridge, Massachusetts, 02139

Published: February 2001

When measuring interfacial tensions, conventional spinning drop tensiometry requires knowledge of the index of refraction of the denser, outer fluid phase, in order to correct for the magnification of the inner, spinning drop. We have modified a spinning drop tensiometer to accommodate a square section sample tube, which does not require this correction. Tests performed using calibration rods are used to demonstrate this fact. The interfacial tension of toluene and water is also measured using this new design. The ensuing results are in agreement with measurements obtained using a conventional circular sample tube. Copyright 2001 Academic Press.