Sensitive high-performance liquid chromatographic assay for motexafin gadolinium and motexafin lutetium in human plasma.

J Chromatogr B Biomed Sci Appl

Program of Molecular Therapeutics and Drug Discovery, University of Pittsburgh Cancer Institute, PA 15213, USA.

Published: December 2000

We present new HPLC methods for the quantitation in human plasma of two investigative metallotexaphyrin agents, motexafin gadolinium (Gd-Tex) and motexafin lutetium (Lu-Tex). Each assay uses: the other texaphyrin analogue as an internal standard; protein precipitation with acetonitrile:methanol (50:50, v/v); an ODS reversed-phase column; an isocratic mobile phase of 100 mM ammonium acetate, pH 4.3:acetonitrile:methanol (59:21:20, v/v/v); and absorbance detection at 470 nm. The Gd-Tex assay has a lower limit of quantitation (LLOQ) of 0.01 microM and is linear between 0.01 and 30 microM. The Lu-Tex assay has an LLOQ of 0.1 microM and is linear between 0.1 and 30 microM. The assays are suited for in vivo preclinical studies and clinical trials because they require minimal amounts of plasma, are sensitive, and involve a 30-mm run time. These assays are important tools for evaluating the potential of Gd-Tex and Lu-Tex as a radiation enhancer and photosensitizer, respectively.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0378-4347(00)00390-xDOI Listing

Publication Analysis

Top Keywords

motexafin gadolinium
8
motexafin lutetium
8
human plasma
8
lu-tex assay
8
001 microm
8
microm linear
8
sensitive high-performance
4
high-performance liquid
4
liquid chromatographic
4
assay
4