Heart rate monitors and abnormal heart rhythm detection.

Arch Physiol Biochem

Laboratoire Performance Motrice, UFR STAPS, Université Blaise Pascal, Aubière Cedex, France.

Published: October 2000

AI Article Synopsis

  • The study evaluates the accuracy of two advanced heart rate monitors, Accurex Plus and Vantage NV, against an older model, PE 4000, in both stable and abnormal heart rhythm scenarios.
  • Results show that the newer monitors were significantly more accurate during stable heart rates, but all models struggled with detecting certain isolated heart rhythm disorders and exhibited a smoothing effect in their measurements.
  • Despite limitations in their ability to detect minor heart rhythm abnormalities, the monitors can still be useful for identifying serious conditions like significant pauses or extreme fluctuations in heart rate.

Article Abstract

We test the accuracy of two third generation heart rate monitors (HRM), Accurex Plus and Vantage NV and compare these data with those of an older monitor, PE 4000. Then we investigate responses of the HRM's to abnormal heart rhythms (HR). Accurex Plus and Vantage NV, and PE 4000 were tested in the laboratory under both stable and transient conditions. The Phantom 320 HR simulator gave 7 stable heart rates, each lasting 3 min. The responses of the HRM's to abnormal heart rhythms were tested with an automatic arrhythmia program lasting 17 min. The new HRM's, Accurex Plus and Vantage NV, were more accurate (89 to 94% and over 98% at +/-3 beats per min) in the stable state than the PE 4000. They all showed a smoothing effect in transient measurements, which was stronger in decelerating HR (1st case) than in accelerating HR (2nd case). - 1st case: Accurex Plus =Vantage NV >PE 4000 -2nd case: PE 4000 >Accurex =Vantage. Most isolated heart rhythm disorders (missing beat, pause < 4 s, SVEA, and SVT) were either not detected or rejected. Signal changes such as changes in QRS (ventricular bigeminy) may cause detection losses, and so modify calculated heart rate. HR signal rising/dropping slopes caused by sudden heart rhythm disorders, was notably attenuated, with a consequent loss of HR accuracy. Although the usefulness of HRM's in detecting HR disorders is limited, they have some value with serious disorders, like pause >4 s, important bradycardia or tachycardia.

Download full-text PDF

Source
http://dx.doi.org/10.1076/apab.108.4.371.4304DOI Listing

Publication Analysis

Top Keywords

heart rate
12
abnormal heart
12
heart rhythm
12
accurex vantage
12
heart
9
rate monitors
8
responses hrm's
8
hrm's abnormal
8
heart rhythms
8
lasting min
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!