Comparing a parallel PERG, automated perimetry, and frequency-doubling thresholds.

Invest Ophthalmol Vis Sci

Centre for Visual Sciences, Research School of Biological Sciences, Australian National University, Canberra. Prince of Wales Hospital, Sydney, Australia.

Published: November 2000

Purpose: A pattern electroretinogram (PERG) simultaneously displaying the frequency-doubling (FD) illusion in nine parts of the visual field was compared with two other methods for ability to detect glaucoma. This multiregion FD PERG (MFP) was compared with results from achromatic automated perimetry and psychophysical tests using FD stimuli.

Methods: MFP data were compared with that from the Humphrey Field Analyser (HFA; Humphrey, San Leandro, CA) 24-2 program. Contrast thresholds were also determined in different visual field locations for FD stimuli. Thin-plate spline methods were used to derive comparisons from the tests, each of which sampled the visual field differently.

Results: Significant correlation with HFA could be obtained, providing seven to nine (of nine) MFP amplitudes were themselves significant. Evidence showed that both the psychophysical tests using FD stimuli and the MFP detect glaucomatous damage not detected by the HFA.

Conclusions: The comparisons between HFA perimetry, the MFP, and FD thresholds indicate that both FD-based tests quantify a form of diffuse loss in early glaucoma as well as the scotomas of later glaucoma.

Download full-text PDF

Source

Publication Analysis

Top Keywords

visual field
12
automated perimetry
8
psychophysical tests
8
mfp
5
comparing parallel
4
parallel perg
4
perg automated
4
perimetry frequency-doubling
4
frequency-doubling thresholds
4
thresholds purpose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!