Assessment of the ion-trap mass spectrometer for routine qualitative and quantitative analysis of drugs of abuse extracted from urine.

J Anal Toxicol

Office of the Armed Forces Medical Examiner, Division of Forensic Toxicology, Armed Forces Institute of Pathology, Rockville, Maryland 20850-3125, USA.

Published: October 2000

The ion-trap mass spectrometer (MS) has been available as a detector for gas chromatography (GC) for nearly two decades. However, it still occupies a minor role in forensic toxicology drug-testing laboratories. Quadrupole MS instruments make up the majority of GC detectors used in drug confirmation. This work addresses the use of these two MS detectors, comparing the ion ratio precision and quantitative accuracy for the analysis of different classes of abused drugs extracted from urine. Urine specimens were prepared at five concentrations each for amphetamine (AMP), methamphetamine (METH), benzoylecgonine (BZE), delta9-carboxy-tetrahydrocannabinol (delta9-THCCOOH), phencyclidine (PCP), morphine (MOR), codeine (COD), and 6-acetylmorphine (6-AM). Concentration ranges for AMP, METH, BZE, delta9-THCCOOH, PCP, MOR, COD, and 6-AM were 50-2500, 50-5000, 15-800, 1.5-65, 1-250, 500-32000, 250-21000, and 1.5-118 ng/mL, respectively. Sample extracts were injected into a GC-quadrupole MS operating in selected ion monitoring (SIM) mode and a GC-ion-trap MS operating in either selected ion storage (SIS) or full scan (FS) mode. Precision was assessed by the evaluation of five ion ratios for n = 15 injections at each concentration using a single-point calibration. Precision measurements for SIM ion ratios provided coefficients of variation (CV) between 2.6 and 9.8% for all drugs. By comparison, the SIS and FS data yielded CV ranges of 4.0-12.8% and 4.0-11.2%, respectively. The total ion ratio failure rates were 0.2% (SIM), 0.7% (SIS), and 1.2% (FS) for the eight drugs analyzed. Overall, the SIS mode produced stable, comparable mean ratios over the concentration ranges examined, but had greater variance within batch runs. Examination of postmortem and quality-control samples produced forensically accurate quantitation by SIS when compared to SIM. Furthermore, sensitivity of FS was equivalent to SIM for all compounds examined except for 6-AM.

Download full-text PDF

Source
http://dx.doi.org/10.1093/jat/24.7.595DOI Listing

Publication Analysis

Top Keywords

ion-trap mass
8
mass spectrometer
8
extracted urine
8
ion ratio
8
concentration ranges
8
operating selected
8
selected ion
8
ion ratios
8
ion
6
sim
5

Similar Publications

Ixeris sonchifolia (IS) has been demonstrated to have beneficial effects on clearing heat and detoxifying, promoting blood circulation and removing blood stasis. However, the protective effects of active fractions and the underlying mechanisms of IS against toxic heat and blood stasis syndrome (THBSS) remain unclear. This study aimed to investigate this.

View Article and Find Full Text PDF

Ammonium bicarbonate buffer system for DNA hybridization and quantification by LC-ESI-MS/MS.

Anal Biochem

March 2025

Research Analytical Biochemistry Laboratory, Department of Chemistry and Biology, Ryerson University, Canada. Electronic address:

High salt buffer may be used for the UV/VIS or radiometric based detection of trihybrid DNA but would contaminate the electrospray mass spectrometer. Colorimetric DNA hybridization assays with the substrate BCIP/NBT reacted with the alkaline phosphatase-streptavidin (APSA) enzyme conjugate that showed a linear range for HIV DNA from 1 pM to 100 pM DNA in presence of high salt concentrations. Ammonia bicarbonate (AMBIC) or ethanolamine resulted in strong DNA hybridization similar to NaCl but was compatible with specific and sensitive mass spectrometry.

View Article and Find Full Text PDF

Additional multipole fields are unavoidable in real quadrupole linear ion traps (QLITs) and play a crucial role in influencing their performance. In this study, the impact of these multipole fields on ion ejection and dynamics in QLITs is exhaustively analyzed using a vectorized Runge-Kutta method and a comprehensive theoretical model of ion vibration involving all the common multipole fields. The comparison of nonlinear resonance under different added multipole fields reveals obvious ion ejection from hexapole and octopole resonances as well as multiple resonance points in most multipole fields.

View Article and Find Full Text PDF

Rationale: To increase the confidentiality level of molecular tags, binary information can be hidden in sequence-defined oligomers using isobaric monomers including a photo-sensitive one, which mass can be modified upon light irradiation. Reading such secured data requires light edition prior to MS/MS sequencing, which was investigated here in a single gas-phase workflow.

Methods: Oligo (phosphodiester)s made of units with either light-sensitive ortho- or light-inert para-nitrobenzyl side-group were electrosprayed as deprotonated oligomers and sampled in an ion trap instrument equipped with various light sources.

View Article and Find Full Text PDF

To evaluate the migration, transformation, and fate of tetrabromobisphenol A (TBBPA) in the environment, the transformation/degradation (T/D) products of TBBPA and byproducts of industrial production should be distinguished. Herein, 7 reported T/D products (R1-R7) and 7 novel byproducts (N1-N7) of TBBPA were identified in industrial-grade TBBPA chemicals by using high-performance liquid chromatography coupled ion trap mass spectrometry and high-resolution mass spectrometry with a suspect screening strategy. The possible formation pathways of these byproducts were attributed to the bromination, debromination, methylation, demethylation, hydroxylation, substitution, and radical coupling reactions of bisphenol A (BPA), BPA impurities, and TBBPA.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!