Purpose: The study compared the setup reliability of 3 patient immobilization systems, a rubber leg cushion, the alpha cradle, and the thermoplastic Hipfix device, in 77 patients with cT1-T3, N0, M0 prostate cancer receiving conformal radiotherapy.

Methods And Materials: Port films were analyzed and compared to simulation films to estimate the setup errors in the three coordinate axes (anterior-posterior, cranial-caudal, medial-lateral). A total vector error was calculated from these shifts.

Results: The Hipfix was found significantly superior to the other two devices in reducing mean setup errors in all axes (p < 0.005). The average field-positioning error with the Hipfix ranged from 1.9 mm to 2.6 mm for all axes, whereas the deviation for the other two systems ranged from 2.7 to 3. 4 mm. Errors greater than 10 mm were virtually eliminated with the Hipfix system. There was a reduction in the mean total vector error in the alpha cradle and Hipfix patient cohorts over time, reflecting improved efficacy as a result of experience.

Conclusion: There was a significant difference in the performance of each immobilization device. The Hipfix was consistently more reliable in reducing setup errors than the other devices.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0360-3016(00)00682-9DOI Listing

Publication Analysis

Top Keywords

setup errors
12
patient immobilization
8
alpha cradle
8
total vector
8
vector error
8
reducing setup
8
hipfix
6
prospective comparison
4
comparison three
4
three systems
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!