Objective: To compare the clinical performance of two intrauterine devices (the frameless GyneFix intrauterine device (IUD) and the TCu380A IUD) during long-term use.
Study Design: This was a randomized comparative study. A total of 157 insertions of the GyneFix IUD were compared with 156 insertions of the TCu380A IUD in the Tianjin Municipal Institute for Family Planning, China. Only parous women were included in the study. Insertions were conducted by the first three authors. The trial extended over a 9-year period, with over 10,000 woman-months of experience.
Results: The study showed that both devices are highly efficacious. No pregnancies occurred with the GyneFix IUD (cumulative pregnancy rate 0.0), compared to one pregnancy with the TCu380A IUD (cumulative pregnancy rate 0.64). The cumulative expulsion rate with the GyneFix IUD was 1.91 (three expulsions occurred, two of which were during the first 6 months), compared to a cumulative expulsion rate of 5.13 (eight expulsions) with the TCu380A (p = 0.1112). The total use-related discontinuation rate (terminations were mainly due to bleeding disturbances) was significantly lower (p = 0.0051) with the GyneFix IUD (4.46) than with the TCu380A IUD (14.10) and resulted in a higher rate of continuation with the GyneFix IUD compared to the TCu380A IUD (92.36 versus 83.97). Neither perforations nor cases of pelvic inflammatory disease were encountered with either device in this study, demonstrating the safety of the anchoring system.
Conclusion: The results of the present clinical trial are in agreement with the findings of other multicenter, randomized, comparative, clinical trials that have investigated these two IUDs in China, as well as with the findings of other large-scale international studies conducted with the frameless IUD.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/13625180008500386 | DOI Listing |
Contraception
June 2024
Contraceptive Development Program, Division of Population Health Research, Eunice Kennedy Shriver National Institute of Child Health and Human Development, Bethesda, MD, USA.
Objective: To evaluate menstrual cup use and intrauterine device (IUD) expulsion.
Study Design: We performed a secondary analysis of a 3-year contraceptive efficacy trial comparing two copper 380 mm IUDs. Investigators randomized participants approximately 1:4 to the TCu380A or NTCu380-Mini IUD.
Contraception
April 2024
Department of Obstetrics and Gynecology, University of Campinas Faculty of Medical Sciences, Campinas, São Paulo, Brazil. Electronic address:
Objectives: This study aimed to compare the uptake, reasons for discontinuation, continuation, and satisfaction of long-acting contraceptive methods (LARC) when offered at no cost in Campinas, Brazil.
Study Design: This was a prospective cohort study conducted at University of Campinas, Campinas, São Paulo, Brazil, including women aged 18 to 48 years, who adopted one of three methods. Participants underwent telephone follow-up every 3 months for up to 24 months after insertion.
J Pediatr Adolesc Gynecol
April 2024
Department of Obstetrics and Gynecology, University of Campinas, Faculty of Medical Sciences, Campinas, SP, Brazil. Electronic address:
Study Objective: To assess clinical performance, bleeding pattern, dysmenorrhea, and satisfaction up to 1 year after placement of 3 types of intrauterine devices (IUDs) (TCu380A, levonorgestrel 52 mg, and levonorgestrel 19.5 mg) in adolescents METHODS: The study was a randomized trial with 318 adolescents allocated to the 3 IUDs. We assessed reasons for removal, continuation, menstrual patterns, dysmenorrhea, and satisfaction.
View Article and Find Full Text PDFEur J Contracept Reprod Health Care
June 2023
Department of Obstetrics and Gynaecology, Faculty of Medical Sciences, University of Campinas (UNICAMP), Campinas, Brazil.
Objective: We compared pain and the ease of insertion of two different types of intrauterine devices (IUDs; the levonorgestrel intrauterine systems 52 mg (LNG-IUS 52 mg) and the copper-bearing IUD [TCu380A]) among nulligravidas and parous women with and without caesarean delivery.
Methods: A cross-sectional study assessed pain referred by women and ease of insertion of the two IUDs models. We applied a visual analogue scale (VAS) to assess pain, while the ease of insertion was defined by healthcare professionals.
Int J Gynaecol Obstet
June 2023
FIGO, Indian College of Obstetricians and Gynaecologists, Purandare Hospital, Mumbai, India.
Background: The intrauterine contraceptive device TCu380Ag when compared with TCu380A at 1 year of use had better acceptability and continuation rates.
Objectives: To study the continuation rate, efficacy, and acceptability of TCu380Ag in three sizes versus TCu380A at 5 years of use.
Methods: A total of 600 women opting for intrauterine contraceptive devices were randomized equally into two groups.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!