A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates. | LitMetric

Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates.

Diabetes Care

Centre Hospitalier Universitaire de Québec, Hospital St. François d'Assise, Laval University, Québec City, Canada.

Published: August 2000

Objective: To evaluate the clinical performance of the Glucometer Elite XL Diabetes Care System in neonatal settings using a multicenter study

Research Design And Methods: A total of 388 blood specimens from 333 neonates were included in the study. A capillary or arterial sample was analyzed for determination of glucose with the Glucometer Elite XL system by an attending trained nurse. Through the same sampling site, a specimen was collected and sent to the laboratory for measurement of plasma glucose, bilirubin, and hematocrit.

Results: The regression analysis between the results of the Glucometer Elite XL system and comparative methods resulted in the following: Glucometer Elite XL meter = 1.01 x laboratory method + 0.02 mmol/l (n = 388). For the 1.1-4.0 mmol/l plasma glucose range, the regression was Glucometer Elite XL meter = 1.07 x laboratory method + 0.12 mmol/l (n = 150). A difference plot indicated a mean bias of 0.04 mmol/l (95% CI -0.01 to 0.10). No relationship was found between meter glucose biases and hematocrit levels (r = 0.10, P = 0.14). Although a statistically significant correlation existed between bilirubin levels and the glucose meter biases (r = 0.14, P = 0.005), the predicted mean biases were of little clinical significance.

Conclusions: The Glucometer Elite XL system showed a good performance when used in neonatal settings.

Download full-text PDF

Source
http://dx.doi.org/10.2337/diacare.23.8.1149DOI Listing

Publication Analysis

Top Keywords

glucometer elite
28
elite meter
12
elite system
12
neonatal settings
8
plasma glucose
8
laboratory method
8
glucometer
7
elite
7
meter
5
glucose
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!