This paper describes systematic methods for measuring and controlling sound levels within a magnetic resonance scanner. The methods are illustrated by application to the acoustic noise generated by a 3 T scanner during echoplanar imaging (EPI). Across five measurement sessions, sound pressure levels at the center of the head gradient coil ranged from 122 to 131 dB SPL [123 to 132 dB(A)]. For protection against damaging noise exposure, UK and US industrial guidelines stipulate that the maximum permitted daily noise dosage is equivalent to 90 dB(A) for 8 hours, where noise dosage is a function of the level of an acoustic signal and the length of exposure to it. Without hearing protection, this equivalent level would be exceeded by less than 5 seconds of exposure to the measured levels of scanner acoustic noise. These findings highlight the importance of noise reduction and hearing protection for those exposed to the acoustic noise generated during EPI.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/1522-2586(200007)12:1<157::aid-jmri17>3.0.co;2-m | DOI Listing |
Nan Fang Yi Ke Da Xue Xue Bao
January 2025
School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China.
Objectives: To explore the synthesis of high-quality CT (sCT) from cone-beam CT (CBCT) using PE-CycleGAN for adaptive radiotherapy (ART) for nasopharyngeal carcinoma.
Methods: A perception-enhanced CycleGAN model "PE-CycleGAN" was proposed, introducing dual-contrast discriminator loss, multi-perceptual generator loss, and improved U-Net structure. CBCT and CT data from 80 nasopharyngeal carcinoma patients were used as the training set, with 7 cases as the test set.
Nan Fang Yi Ke Da Xue Xue Bao
January 2025
School of Mathematics and Computer Science, Gannan Normal University, Ganzhou 341000, China.
Objectives: We present a new low-dose CT reconstruction method using sub-pixel and anisotropic diffusion.
Methods: The sub-pixel intensity values and their second-order differences were obtained using linear interpolation techniques, and the new gradient information was then embedded into an anisotropic diffusion process, which was introduced into a penalty-weighted least squares model to reduce the noise in low-dose CT projection data. The high-quality CT image was finally reconstructed using the classical filtered back-projection (FBP) algorithm from the estimated data.
ArXiv
December 2024
Department of Radiology, Mayo Clinic College of Medicine and Science, Rochester, MN, USA.
Ultrasound localization microscopy (ULM) enables microvascular imaging at spatial resolutions beyond the acoustic diffraction limit, offering significant clinical potentials. However, ULM performance relies heavily on microbubble (MB) signal sparsity, the number of detected MBs, and signal-to-noise ratio (SNR), all of which vary in clinical scenarios involving bolus MB injections. These sources of variations underscore the need to optimize MB dosage, data acquisition timing, and imaging settings in order to standardize and optimize ULM of microvasculature.
View Article and Find Full Text PDFPhys Med Biol
January 2025
School of Mathematical Sciences, Capital Normal University, Beijing 100048, People's Republic of China.
Low-dose computed tomography (LDCT) has gained significant attention in hospitals and clinics as a popular imaging modality for reducing the risk of x-ray radiation. However, reconstructed LDCT images often suffer from undesired noise and artifacts, which can negatively impact diagnostic accuracy. This study aims to develop a novel approach to improve LDCT imaging performance.
View Article and Find Full Text PDFPhys Med Biol
January 2025
Department of Radiology, Stanford University, Stanford, CA 94305, United States of America.
. Radiation dose and diagnostic image quality are opposing constraints in x-ray computed tomography (CT). Conventional methods do not fully account for organ-level radiation dose and noise when considering radiation risk and clinical task.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!