The evaluation of cardiac troponin I (cTnI) on the Dimension RxL-HM analyzer is presented. The one-step enzyme immunoassay is based on two cTnI specific monoclonal antibodies. Performed on a separate module of the analyzer, assay-time is 17 minutes. Using as criterion a between-run impression CV <20% the functional limit of detection was set at 0.1 microg/l. Cutoff level for minor myocardial damage of 0.1 microg/l was found. In Duchenne's dystrophy, patients showed increased cardiac Troponin T (cTnT) but no increased cTnI. In patients with a history of coronary heart disease undergoing chronic hemodialysis, cTnT and cTnI were increased. In different patients with submassive pulmonary embolism, increased cTnI was determined. In coronary artery bypass surgery without perioperative myocardial infarction, patients with extracorporeal circulation showed significantly higher cTnI at 24 h after surgery than those with minimal cardiac surgery. In patients with unstable angina, increased cTnI was found more often than on Stratus analyzer. In conclusion, the new assay is a very sensitive cTnI assay, fast and easy to perform in parallel to enzyme and substrate assays.

Download full-text PDF

Source
http://dx.doi.org/10.1515/CCLM.2000.051DOI Listing

Publication Analysis

Top Keywords

dimension rxl-hm
8
analytical clinical
4
clinical evaluation
4
evaluation troponin
4
troponin determination
4
determination dimension
4
rxl-hm evaluation
4
evaluation cardiac
4
cardiac troponin
4
troponin ctni
4

Similar Publications

Objective: We compared the results of 2 immunoassay systems (affinity column-mediated immunoassay [ACMIA] and microparticle enzyme immunoassay [MEIA]), regarding hematologic and biochemical values at 2 weeks after liver transplantation.

Methods: We obtained 256 blood samples from 35 patients, at 2 weeks after liver transplantation, excluding those from patients who were treated with interacting medications or renal replacement therapy. We also excluded the early mortality cases within 2 weeks of liver transplantation.

View Article and Find Full Text PDF

Background: Tacrolimus is a potent immunosuppressive drug used in organ transplantation. Because of its substantial toxic effects, narrow therapeutic index, and interindividual pharmacokinetic variability, therapeutic drug monitoring of whole-blood tacrolimus concentrations has been recommended. We investigated the comparability of the results of 2 immunoassay systems, affinity column-mediated immunoassay (ACMIA) and microparticle enzyme immunoassay (MEIA), comparing differences in the tacrolimus concentrations measured by the 2 methods in relation to the hematologic and biochemical values of hepatic and renal functions.

View Article and Find Full Text PDF

[Evaluation of the analytical performance characteristics of NT-proBNP immunoassay on dimension RxL-HM (Dade Behring)].

Ann Biol Clin (Paris)

October 2005

Laboratoire de biochimie, Hôpital européen Georges Pompidou, Paris.

Background: Plasma B-type natriuretic peptide (BNP) and N terminal prohormone (NT-proBNP) are promising markers for the diagnosis, prognosis and follow up of heart failure. Elevated levels have been associated with adverse long-term outcome in patients with acute coronary syndromes.

Method: The analytical performance of a new automated NT-proBNP assay on Dimension-HM system (Dade Behring) and comparison with the results obtained with the Roche Diagnostics NT-proBNP assays using the Elecsys instrument have been evaluated according to the Valtec protocol.

View Article and Find Full Text PDF

[Evaluation of the heterogeneous immunoassay (ACMIA) for the measurement of blood cyclosporin on the Behring dimension RXL clinical chemistry analyzer].

Ann Biol Clin (Paris)

March 2004

Laboratoire de pharmacologie et toxicologie, CHU Henri Mondor, 51, avenue du Maréchal de Lattre de Tassigny, 94010 Créteil.

We propose an evaluation of a new heterogeneous immunoassay of cyclosporin on RXL HM Dimension (Dade Behring) for therapeutic cyclosporin monitoring in whole-blood patients transplant. The pretreatment step is performed automatically into the apparatus while it is a manual step with EMIT. Linearity, intra- and inter-day precision, limit of quantification, precision and accuracy of dilution steps and stability into the equipment were studied.

View Article and Find Full Text PDF

[Evaluation of Emit netilmicin and amikacin assays on Dimension RXL HM (Dade Behring) open channels].

Ann Biol Clin (Paris)

February 2003

Laboratoire de biochimie, bâtiment 3B, Centre hospitalier Lyon-Sud, Chemin du grand Revoyet, 69495 Pierre-Bénite Cedex, France.

The aim of this study was to evaluate and to validate an enzyme immunoassay in homogeneous phase for netilmicin and amikacin, adapted on the Dimension RXL HM (Dade Behring) machine. The results were compared with those obtained with automated polarization of fluorescence immunoassay using TDx FLx (Abbott). The protocol of the study and the analytical criteria were inspired by the protocol Valtec version 2002 recommended by the French Society of Clinical Biology (SFBC).