Unlabelled: A new "all in one" sensing device was developed for continuous transtracheal intraoperative monitoring and in situ detection of the recurrent laryngeal nerve (RLN) during thyroid surgery.

Patients And Methods: The new system is based on a double-balloon endotracheal tube with integrated atraumatic stimulating and tracing electrodes. The recurrent laryngeal nerve is stimulated transtracheally and compound action potentials are recorded from the laryngeal muscles. Fifty-five patients were introduced into a phase-one clinical trial. Thirty-five patients with primary thyroid operations, 20 patients with reoperations, 10 of whom had neck dissections. All patients were evaluated laryngoscopically and phoniatrically by an ENT specialist before and after surgery.

Results: Compound muscle action potentials were recorded continuously during the whole operation and responded sensitively to tension and pressure to the nerve. There were no accidental permanent RLN palsies.

Conclusion: The new system offers five advantages: (1) it is atraumatic; (2) it is easy to use; (3) it can monitor continuously with an audio feedback to the surgeon; (4) it works outside the operation field; and (5) it is highly sensitive, even indicating reversible irritation to the nerve.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s001040051101DOI Listing

Publication Analysis

Top Keywords

recurrent laryngeal
8
laryngeal nerve
8
action potentials
8
potentials recorded
8
nerve
5
[first continuous
4
continuous nerve
4
nerve monitoring
4
monitoring thyroid
4
thyroid gland
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!