Quality assurance by systematic in vivo dosimetry: results on a large cohort of patients.

Radiother Oncol

Servizio di Fisica Sanitaria, H.S. Raffaele, Via Olgettina 60, 20132, Milano, Italy.

Published: July 2000

Background: In vivo dosimetry is widely considered to be an important tool for quality assurance in external radiotherapy.

Introduction: In this study we report on our experience over more than 4 years in systematic in vivo dosimetry with diodes.

Materials And Methods: From November '94 an in vivo entrance dosimetry check was performed for every new patient irradiated at one of our treatment units (Linac 6/100, 6 MV X-rays). Diodes were calibrated in terms of entrance dose; appropriate correction factors had been previously assessed (taking SSDs, field width, wedge, oblique incidence and blocking tray into account) and were individually applied to in vivo diode readings. The in vivo measured entrance dose was compared with the expected one, with a 5% action level; if a larger deviation was found, all treatment parameters were verified, and the in vivo dosimetry check was repeated. During the period November '94-May '99, 2824 measurements on 1433 patients were collected.

Results: Nine out of 1433 (0.63%) serious systematic errors (leading to a 5% or more on the delivered dose to the PTV) were detected by in vivo dosimetry; four out of nine would produce a 10% or more error if not detected. The rate of serious systematic errors detected by an independent check of treatment chart and MU calculation was found to be 1.5%, showing that less than 1/3 of the errors escapes this check. One hundred and twelve out of 1433 (7.8%) patients had more than one check: the rate of second checks was significantly higher for breast patients (31/250, 12.4%) against non-breast patients (81/1183, 6.8%, P=0.003). A number of patients demonstrated a persistent relatively large error even after two or more checks. For almost all patients the cause of the deviation was assessed; the most frequent cause was the difficulty in correctly positioning the patient and/or the diode. When analyzing the distribution of the deviations between measured and expected entrance doses (excluding first checks in the case of repetition of the in vivo dosimetry control) the mean deviation was 0.4% with a standard deviation equal to 3.0%. The rates of deviations larger than 5 and 7% were 9.9 and 2.6%, respectively. When considering the same data taking the average deviation in the case of opposed beams, the SD became 2.6% and the rates of deviations larger than 5 and 7%, respectively, 5.2 and 0.8%. When dividing the beams according to their orientation, significantly higher rates of large deviations (>5 and 7%) were found for oblique and posterior-anterior (PA) fields against lateral and anterior-posterior (AP) fields (P<0.05). Similarly, higher rates of large deviations were found for wedged fields against unwedged fields (P<0.03) and for blocked fields against unblocked fields (P<0.01). When dividing the data according to the anatomical district, accuracy was worse for breast (mean deviation 0.1%, 1 SD: 3.5%) and neck AP-PA fields (mean deviation 1%, 1 SD: 3,4%). Better accuracy was found for vertebrae (0.1%, 1 SD 2. 1%) and brain patients (-0.7%, 1 SD: 2.6%). During the considered period, in vivo dosimetry was also able to promptly detect a systematic error caused by a wrong resetting of the simulator height couch indicator, with a consequent error in the estimate of patient thickness of about 4 cm.

Conclusions: In our experience, systematic in vivo dosimetry demonstrated to be a valid tool for quality assurance, both in detecting systematic errors which may escape the data transfer/MU calculation check and in giving an effective way of estimating the accuracy of treatment delivery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0167-8140(00)00195-xDOI Listing

Publication Analysis

Top Keywords

vivo dosimetry
24
vivo
9
quality assurance
8
systematic vivo
8
dosimetry check
8
entrance dose
8
serious systematic
8
systematic errors
8
rates deviations
8
deviations larger
8

Similar Publications

Brachytherapy is a key treatment for gynaecological malignancies, delivering high doses to the tumour volume whilst sparing nearby normal tissues due to its steep dose gradient. Accuracy is imperative as small shifts can lead to clinically significant under- or over-dosing of the target volume or organs at risk (OARs), respectively. Independent verification of dose delivered during brachytherapy is not routinely performed but it is important to identify gross errors and define action thresholds to guide inter-fraction treatment decisions.

View Article and Find Full Text PDF

Background: The use of in-vivo dosimetry is a long-standing but also labor-intensive component of risk-level assessment for patients with implanted devices. A calculation-only approach, using treatment planning system (TPS)-calculated doses along with imaging doses estimates when relevant, has the potential to streamline the physics workflow without negatively impacting patient safety.

Purpose: To evaluate the feasibility of using a calculation-only approach for risk level assessment for patients with implanted electronic medical devices.

View Article and Find Full Text PDF

Risk assessment and quality management in AIO based on CT-linac for nasopharyngeal carcinoma: An improved FMEA and FTA approach.

Med Phys

January 2025

State Key Laboratory of Oncology in South China, Guangdong Key Laboratory of Nasopharyngeal Carcinoma Diagnosis and Therapy, Guangdong Provincial Clinical Research Center for Cancer, Sun Yat-sen University Cancer Center, Guangzhou, P. R. China.

Article Synopsis
  • AIO radiotherapy integrates all conventional steps into one device, improving efficiency and reducing errors for cancer patients while facing challenges like software complexity and AI reliance.
  • A risk assessment using FMEA and FTA was conducted to evaluate the quality management measures for nasopharyngeal carcinoma treatment.
  • Results showed 86 failure modes identified, with risk priority numbers indicating varying levels of risk, emphasizing the importance of implementing quality management measures for safety.
View Article and Find Full Text PDF

[18F]Tetrafluoroborate: a new NIS PET/CT radiopharmaceutical. An overview focused on differentiated thyroid cancer.

Eur Thyroid J

January 2025

G Treglia, Repubblica e Cantone Ticino Ente Ospedaliero Cantonale, Bellinzona, Switzerland.

Background: In relapsing differentiated thyroid cancer (DTC), the in vivo evaluation of natrium-iodine symporter (NIS) expression is pivotal in the therapeutic planning and is achieved by [131/123I]Iodine whole-body scan. However, these approaches have low sensitivity due to the low sensitivity due to the low resolution of SPECT. [18F]Tetrafluoroborate (TFB) has been proposed as a viable alternative, which could outperform [131/123I]Iodine scans owing to the superior PET resolution.

View Article and Find Full Text PDF

Purpose Low-dose total skin electron beam therapy (LD-TSEBT) has recently gained popularity in treating mycosis fungoides (MF) due to its reduced toxicity and favorable response rates. Combining accelerated LD-TSEBT with the modified Stanford technique (mST), a condensed cycling approach, offers a promising and convenient option. However, in vivo dosimetry data confirming the effectiveness of this approach is limited.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!