Background And Objective: The commercially available optical coherence tomography (OCT) scanner displays images in a pre-set window regardless of the projected scan length on the retinal surface. The aim of this study was to demonstrate the true dimensions of proportion-corrected OCT images and the additional information present in grayscale images.

Materials And Methods: OCT raw data were exported to an IBM-compatible PC and processed to show grayscale and proportion-corrected images using an automated software of our own design.

Results: Eyes with cystoid macular edema and retinal pigment detachment were analyzed. Grayscale images showed a finer gradation of signal reflectance. Scan lengths of 2, 4, 6, and 8 mm on the retinal surface showed different qualitative appearances using proportion-corrected software from the printed or on-screen images.

Conclusions: Grayscale OCT images can be used to demonstrate additional information not present in false-color images. The disparity between the standard OCT image format and proportion-corrected images emphasizes the need for quantitative rather than qualitative evaluation of retinal dimensions and internal reflectance.

Download full-text PDF

Source

Publication Analysis

Top Keywords

grayscale proportion-corrected
8
optical coherence
8
coherence tomography
8
images
8
retinal surface
8
oct images
8
proportion-corrected images
8
grayscale
5
oct
5
proportion-corrected optical
4

Similar Publications

Grayscale and proportion-corrected optical coherence tomography images.

Ophthalmic Surg Lasers

September 2000

Department of Ophthalmology, The New York Eye and Ear Infirmary, NY 10003, USA.

Background And Objective: The commercially available optical coherence tomography (OCT) scanner displays images in a pre-set window regardless of the projected scan length on the retinal surface. The aim of this study was to demonstrate the true dimensions of proportion-corrected OCT images and the additional information present in grayscale images.

Materials And Methods: OCT raw data were exported to an IBM-compatible PC and processed to show grayscale and proportion-corrected images using an automated software of our own design.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!