Purpose: We describe a method of performing salvage prostate brachytherapy on patients whose initial implant was suboptimal. This technique uses an image correlation algorithm only previously used to fuse postimplant magnetic resonance and computed tomographic (CT) images. Here, the initial postimplant CT and the second preimplant volume study are coregistered to plan delivery of the salvage implant.
Materials And Methods: Two early-stage patients had salvage implants performed with this technique, in which only a limited number of sources were visible on the ultrasound images. The dosimetric results of the first implant were displayed on the preplan generated for the second procedure. The planned total dose then was visualized prior to salvage implant.
Results: The implants were performed without complication. Rectum and urethra doses remained acceptable. In each case, the improvement in coverage of the gland was dramatic (V80 coverage improved from 65.2% and 47.3% to 93.1% and 92.2%, respectively), precluding the need for further intervention.
Conclusions: Coregistration of the postimplant CT scan to an ultrasound volume study can be quantifiably and reliably performed. The resulting image set can be used to guide needle placement during a second salvage implant to achieve much improved dosimetric coverage of the gland.