A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy.

Pediatr Radiol

Department of Diagnostic Radiology, Oregon Health Science University, Doernbecher Children's Hospital, Portland 97201-3098, USA.

Published: April 2000

Background: Traditionally, pediatric radiologists have been advocates of fluoroscopy systems that provide diagnostic images at the lowest possible radiation dose to the pediatric patient. Manufacturers of fluoroscopic equipment vary as to their claims of "low radiation" exposures.

Objectives: To obtain comparative data on radiation exposure and image quality from four pediatric hospitals, across variants of fluoroscopic equipment (such as pulsed versus continuous fluoroscopy).

Materials And Methods: Images were acquired from phantoms that simulated the size of a 3-year-old child. Phantom results, both stationary and rotating dynamic, were evaluated for radiation exposure and for image resolution of high- and low-contrast objects.

Results: Radiation exposure from the four fluoro units varied widely; the lowest-dose selectable fluoro mode produced exposures varying between 34 and 590 mrads/min among the four fluoro units, and the highest-dose selectable fluoro mode produced 540-2,230 mrads/min. The lowest radiation exposures were produced by pulsed fluoro units, and the very lowest radiation exposure was produced by a fluoroscope that had been especially optimized for pediatric imaging. There was only a small variation in image quality among the hospitals for visualization of stationary objects. A wide variability was noted for detection of objects on the moving phantom.

Conclusions: The variability in the number of detected objects was considerably smaller than the variability in radiation exposure. Pulsed fluoroscopy provides improved resolution for moving objects. Optimization of one hospital's fluoroscope especially for pediatric imaging produced the best ratio of image quality to radiation exposure.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s002470050729DOI Listing

Publication Analysis

Top Keywords

radiation exposure
28
image quality
16
exposure image
12
lowest radiation
12
fluoro units
12
radiation
9
quality pediatric
8
fluoroscopic equipment
8
selectable fluoro
8
fluoro mode
8

Similar Publications

Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.

Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.

View Article and Find Full Text PDF

Whilst the world sees the tremendous growth of mobile phone technology, radiofrequency electromagnetic radiation (RF-EMR) induced possible health effects have emerged as a topic of recent day debate. The current study is designed to test the hypothesis that chronic 900MHz radiation exposure would potentially dysregulate the stress response system (HPA axis) in vivo, via, its non-thermal mechanisms, leading to alterations in the microarchitecture of the adrenal gland, vulnerable brain regions such as the hippocampus which may results in altered behaviours in rats. Male albino Wistar rats aged four weeks, weighing 50-60g were subjected to 900MHz radiation from a cellphone for four weeks at a rate of one hour per day.

View Article and Find Full Text PDF

Investigation of the Ocular Response and Corneal Damage Threshold of Exposure to 28 GHz Quasi-millimeter Wave Exposure.

Health Phys

January 2025

Division of Vision Research for Environmental Health, Medical Research Institute and Department of Ophthalmology, Kanazawa Medical University, Kahoku, Japan.

Electromagnetic radiation energy at millimeter wave frequencies, typically 30 GHz to 300 GHz, is ubiquitously used in society in devices for telecommunications; radar and imaging systems for vehicle collision avoidance, security screening, and medical equipment; scientific research tools for spectroscopy; industrial applications for non-destructive testing and precise measurement; and military and defense applications. Understanding the biological effects of this technology is essential. We have been investigating ocular responses and damage thresholds comparing various frequencies using rabbit eyes and dedicated experimental apparatus.

View Article and Find Full Text PDF

Pharmacokinetic and Metabolomic Studies with BBT-059 in Nonhuman Primates Exposed to Total-Body Gamma Radiation.

Radiat Res

December 2024

Division of Radioprotectants, Department of Pharmacology and Molecular Therapeutics, F. Edward Hébert School of Medicine, Uniformed Services University of the Health Sciences, Bethesda, Maryland.

BBT-059 is a long-acting PEGylated interleukin-11 analog that has been shown to have hematopoiesis-promoting and anti-apoptotic attributes, and is being studied as a radiation countermeasure for the hematopoietic acute radiation syndrome (H-ARS). This potential countermeasure has been demonstrated to enhance survival in irradiated mice. To investigate the toxicity and safety profile of this agent, 14 nonhuman primates (NHPs, rhesus macaques) were administered two different doses of BBT-059 subcutaneously 24 h after 4 Gy total-body irradiation and were monitored for the next 60 days postirradiation.

View Article and Find Full Text PDF

Background: Although ultraviolet radiation is the most important risk factor for skin cancer in general, its role in disease control after the diagnosis of melanoma remains to be elucidated.

Objectives: To analyse the association between survival and self-assessed sun protection behaviour after a melanoma diagnosis.

Methods: This was a questionnaire-based pilot study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!