Purpose: To study the variation of computed tomography (CT) number from a simulator-based scanner and the effect of this variation on photon-dose calculations.

Method And Materials: CT images of a cylindrical phantom with multiple inserts were obtained using a commercially-available simulator-CT (Ximatron: Varian, Palo Alto, CA). The linear correlation coefficient and Chi-square methods were used to determine the X-ray effective energy in a phantom. CT numbers in Hounsfield units (HU) were measured as a function of phantom size, orientation, field of view (FOV), distance from the center, and time for various inserts. The change of dose calculations due to the CT number variations was then determined using the equivalent path-length (EPL) and collapsed cone convolution methods.

Results And Discussion: A significant beam-hardening effect was observed for the simulator-CT. Consequently, the CT number from the sim-CT was more sensitive to the size of the phantom than those from a conventional CT. The sim-CT number is not sensitive to the locations within the phantom and is stable over a 6-week period. It is important to use the proper FOV for sim-CT studies; scanning a small polystyrene phantom using a large FOV may result in an increase of l20 HU in CT number at the center of the field. However, the dose-calculation variations, due to the CT number uncertainty, do not exceed 2-3% for 6-18 MV photon beams.

Conclusion: The simulator CT images were acquired with patients in the treatment position, and these CT numbers are useful for CT-based dose calculations.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0167-8140(00)00159-6DOI Listing

Publication Analysis

Top Keywords

dose calculations
12
computed tomography
8
tomography number
8
number
7
phantom
6
applications simulator
4
simulator computed
4
number photon
4
photon dose
4
calculations radiotherapy
4

Similar Publications

Purpose: Mitoxantrone (MTX) is largely restricted in clinical usage due to its significant cardiotoxicity. Multiple studies have shown that an imbalance in the gut-heart axis plays an important role in the development of cardiovascular disease (CVD). We aim to explore the possible correlations between gut microbiota (GM) compositions and cardiometabolic (CM) disorder in MTX-triggered cardiotoxicity mice.

View Article and Find Full Text PDF

The quality of CT images obtained from hepatocellular carcinoma (HCC) patients is complex, affecting diagnostic accuracy, precision, and radiation dose assessment due to increased exposure risks. The study evaluated image quality qualitatively and quantitatively by comparing quality levels with an effective radiation dose to ensure acceptable quality accuracy. This study retrospectively reviewed 100 known HCC patients (Li-RADS-5) who underwent multidetector computed tomography (MDCT) multiphasic scans for follow-up of their health condition between January and October 2023.

View Article and Find Full Text PDF

The therapeutic failure of infliximab therapy remains a challenge in patients with inflammatory bowel disease (IBD), and dose optimization is often required. Accelerated or intensified regimes showed value in treating patients in the acute setting with high CRP or low albumin levels, which are suggested by recent guidelines; however, evidence is weak. Therapeutic drug monitoring (TDM), with anti-tumor necrosis factor-alpha (TNF-α) trough levels and antibodies, showed value during maintenance therapy, but not in induction and can guide clinical decisions in patients that might be undertreated with the standard dosing regimen.

View Article and Find Full Text PDF

The determinants of varicella-zoster virus (VZV)-associated central nervous system (CNS) infection have not been fully elucidated. This study aimed to investigate the incidence and risk factors, including immunosuppression, for different manifestations of VZV-associated CNS infection. Patient registers were used to include adults diagnosed with VZV-associated CNS infections between 2010 and 2019 in Sweden.

View Article and Find Full Text PDF

Background: The use of iodinated contrast-enhancing agents in computed tomography (CT) improves the visualization of relevant structures for radiotherapy treatment planning (RTP). However, it can lead to dose calculation errors by incorrectly converting a CT number to electron density.

Purpose: This study aimed to propose an algorithm for deriving virtual non-contrast (VNC) electron density from dual-energy CT (DECT) data.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!