Quality Determining Parameters of Balloon Angioplasty Catheters.

J Invasive Cardiol

Rostock University, Institute of Biomedical Engineering, P.O. Box 10 08 88, 18055, Rostock, Germany.

Published: April 1996

AI Article Synopsis

  • The study evaluated various balloon angioplasty catheters for stenosis passability using a parameter test unit, allowing a fair comparison of brands.
  • Both over-the-wire and monorail catheter systems demonstrated similar passage capabilities, with certain brands showing consistent performance regardless of repeated inflation.
  • The findings emphasize the importance of material quality in catheter design and suggest careful consideration of balloon durability when discussing catheter reuse.

Article Abstract

The investigations carried out on balloon angioplasty catheters using a parameter test unit allow an objective comparison and a qualitative assessment of catheter brands with respect to their stenosis passability. The design of the test unit allows the measure of other parameters important for clinical practice, such as pushability and trackability. Altogether 8 over-the-wire and 10 monorail-catheters from 7 different manufacturers were investigated by our study. Between modern over-the-wire and monorail systems there exists no fundamental difference in passage capability. In some brands of catheter the stenosis model shows no increase in the amount of thrust required, even after repeated inflation; this is accountable to the good folding properties of the balloon. Angioplasty balloons made of less overstretchable or thinner material yielded the best results. The fact that changes in balloon quality may result from pre-inflation should carry weight in discussions concerning the reuse of catheters. The measurement results are useful for the correct selection of a catheter appropriate to the therapy task at hand.

Download full-text PDF

Source

Publication Analysis

Top Keywords

balloon angioplasty
12
angioplasty catheters
8
test unit
8
quality determining
4
determining parameters
4
balloon
4
parameters balloon
4
catheters investigations
4
investigations carried
4
carried balloon
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!