The Falconª Eccentric Coronary Balloon Angioplasty Catheter: A Randomized Safety and Efficacy Study.

J Invasive Cardiol

Consultant Cardiologist, Department of Cardiology, Warwick Hospital, Lakin Road, Warwick, CV34 5BW, United Kingdom.

Published: October 1997

Percutaneous coronary angioplasty (PTCA) is usually performed using concentric shaped balloon catheters with the guidewire passing through the center of the shaft. The Falconª balloon catheter features a guide wire lumen on the outside of the balloon so that an eccentric balloon catheter profile is obtained concentrating the dilating force on the wire supported side, allowing lower inflation pressures and potentially causing less vessel injury. The aim of this study was to evaluate the safety and efficacy of this new balloon catheter in patients with stable and unstable angina. In 95 prospectively randomized patients, 57 lesions were dilated with a concentric balloon and 51 with the eccentric balloon. Technical success in the two groups was similar (73.3% vs. 74.5% control vs. Falcon respectively). Procedural success was 96.5% vs. 96.1% in the control and Falcon groups respectively. The mean increase in minimum luminal diameter (MLD) was 1.01 +/- 0.41 mm in the control vs 0.85 +/- 0.45 mm in Falcon (p = 0.053). There was an increase in type A dissections in the Falcon group 18 (36.75%) vs. 10 (19.23%) in the control group (p = 0.07) with no difference in stent implantation, myocardial infarction, CABG or death between the two groups. All patients with a technically successful PTCA were followed up. Seventeen (43.6%) in the control and 11 (32.4%) in the Falcon had repeat coronary angiography (p = 0.38), 12 (30.8%) vs. 7 (20.6%) had repeat PTCA (p = 0.37) and time to PTCA was 116 +/- 70 days vs. 154 +/- 103 days respectively (p = 0.36). The Falcon performed technically as well as the concentric balloons. Despite a smaller MLD and increase in Type A dissections there was no associated increase in complications or reintervention for restenosis. Further investigation is required to evaluate the role of this mechanism of dilatation in restenosis.

Download full-text PDF

Source

Publication Analysis

Top Keywords

balloon catheter
12
balloon
8
safety efficacy
8
balloon eccentric
8
eccentric balloon
8
control falcon
8
increase type
8
type dissections
8
falcon
6
control
5

Similar Publications

Background: Several studies have demonstrated a notable increase in the incidence of atrial arrhythmias among individuals with atrial septal defect (ASD) occluder. Although the sequential dilation technique has been proposed as the mainstream technique for transseptal puncture with ASD occluder, it is associated with substantial risks and technical difficulties.

Case Summary: We report a patient who underwent catheter ablation for atrial fibrillation and had a large ASD occluder.

View Article and Find Full Text PDF

Coronary Events After Transcatheter Aortic Valve Replacement: Insights From the France TAVI Registry.

JACC Cardiovasc Interv

January 2025

Institut Cardiovasculaire Paris-Sud, Hôpital Privé Jacques Cartier, Ramsay-Santé, Massy, France. Electronic address:

Background: The prevalence of coronary artery disease in patients undergoing transcatheter aortic valve replacement (TAVR) is high. Treatment of a coronary events (CE) after TAVR can be technically challenging.

Objectives: The authors sought to assess the incidence and prognostic impact of CE after TAVR.

View Article and Find Full Text PDF

A 69-year-old man with chest pain was diagnosed with acute type B aortic dissection with the entry tear located at distal arch and a distal aortic arch aneurysm. Therefore, we performed debranching thoracic endovascular aortic repair 2 weeks after type B aortic dissection onset. First, the graft was anastomosed to bilateral axillary arteries.

View Article and Find Full Text PDF

Background: An effectively functioning arteriovenous fistula (AVF) is vital for end-stage renal disease patients. This study aims to evaluate the effects of Fogarty balloon catheter dilation on creating an effectively functioning AVF.

Methods: This retrospective cohort study was conducted at our clinic between 2020 and 2022.

View Article and Find Full Text PDF

Background: This study aimed to investigate the clinical outcomes of percutaneous transluminal angioplasty (PTA) in patients undergoing hemodialysis with different types of superior vena cava obstruction (SVCO) lesions.

Methods: This retrospective observational study recruited patients undergoing hemodialysis with SVCO and analyzed the clinical characteristics of SVCO. Patency rates were collected for patients treated with PTA and were assessed using the t-test, U-test, log-rank test and survival analyses such as the Kaplan-Meier method.

View Article and Find Full Text PDF