Purpose: To investigate the clinical importance and feasibility of a 3-D portal image analysis method in comparison with a standard 2-D portal image analysis method for pelvic irradiation techniques.

Methods And Materials: In this study, images of 30 patients who were treated for prostate cancer were used. A total of 837 imaged fields were analyzed by a single technologist, using automatic 2-D and 3-D techniques independently. Standard deviations (SDs) of the random, systematic, and overall variations, and the overall mean were calculated for the resulting data sets (2-D and 3-D), in the three principal directions (left-right [L-R], cranial-caudal [C-C], anterior-posterior [A-P]). The 3-D analysis included rotations as well. For the translational differences between the three data sets, the overall SD and overall mean were computed. The influence of out-of-plane rotations on the 2-D registration accuracy was determined by analyzing the difference between the 2-D and 3-D translation data as function of rotations. To assess the reliability of the 2-D and 3-D methods, the number of times the automatic match was manually adjusted was counted. Finally, an estimate of the workload was made.

Results: The SDs of the random and systematic components of the rotations around the three orthogonal axes were 1. 1 (L-R), 0.6 (C-C), 0.5 (A-P) and 0.9 (L-R), 0.6 (C-C), 0.8 (A-P) degrees, respectively. The overall mean rotation around the L-R axis was 0.7 degrees, which deviated significantly from zero. Translational setup errors were comparable for 2-D and 3-D analysis (ranging from 1.4 to 2.2 mm SD and from 1.5 to 2.5 mm SD, respectively). The variation of the difference between the 2-D and 3-D translation data increased from 1.1 mm (SD) for zero rotations to 2.7 mm (SD) for out-of-plane rotations of 3 degrees, due to a reduced 2-D registration accuracy for large rotations. The number of times the analysis was not considered acceptable and was manually adjusted was 44% for the 2-D analysis, and 6% for the 3-D analysis.

Conclusion: True 3-D analysis of setup errors for a group of 30 patients with prostate cancer demonstrated that setup rotations are rather small. The deformation of the projected anatomy in portal images caused by out-of-plane rotations leads to a reduced 2-D registration accuracy. For rotations larger than 3 degrees this effect can be quite pronounced, making 3-D registration the preferred method. Furthermore, the automatic 3-D registration has a higher success rate, most likely because this technique uses more information compared to the 2-D method.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0360-3016(99)00468-xDOI Listing

Publication Analysis

Top Keywords

2-d 3-d
28
3-d analysis
16
3-d
14
2-d
13
portal image
12
image analysis
12
prostate cancer
12
out-of-plane rotations
12
2-d registration
12
registration accuracy
12

Similar Publications

One of the main drivers of fibrotic diseases is epithelial-mesenchymal transition (EMT): a transdifferentiation process in which cells undergo a phenotypic change from an epithelial state to a pro-migratory state. The cytokine transforming growth factor-β1 (TGF-β1) has been previously shown to regulate EMT. TGF-β1 binds to fibronectin (FN) fibrils, which are the primary extracellular matrix (ECM) component in renal fibrosis.

View Article and Find Full Text PDF

Peficitinib suppresses diffuse-type tenosynovial giant cell tumor by targeting TYK2 and JAK/STAT signaling.

Sci China Life Sci

January 2025

Institute of Cardiovascular Sciences, School of Basic Medical Sciences, Peking University Health Science Center; Department of Cardiology and Institute of Vascular Medicine, Peking University Third Hospital; State Key Laboratory of Vascular Homeostasis and Remodeling, Peking University, Beijing, 100191, China.

Diffuse-type tenosynovial giant cell tumor (dTGCT) is a destructive but rare benign proliferative synovial neoplasm. Although surgery is currently the main treatment modality for dTGCT, the recurrence risk is up to 50%. Therefore, there is a great need for effective drugs against dTGCT with minor side effects.

View Article and Find Full Text PDF

Rapid High-Sensitivity Analysis of Methane Clumped Isotopes (ΔCHD and ΔCHD) Using Mid-Infrared Laser Spectroscopy.

Anal Chem

January 2025

Laboratory for Air Pollution/Environmental Technology, Empa, 8600 Dübendorf, Switzerland.

Mid-infrared laser absorption spectroscopy enables rapid and nondestructive analysis of methane clumped isotopes. However, current analytical methods require a sample size of 20 mL STP (0.82 mmol) of pure CH gas, which significantly limits its application to natural samples.

View Article and Find Full Text PDF

The (3+1)-dimensional mKdV-ZK model is an important framework for studying the dynamic behavior of waves in mathematical physics. The goal of this study is to look into more generic travelling wave solutions (TWSs) for the generalized ion-acoustic scenario in three dimensions. These solutions exhibit a combination of rational, trigonometric, hyperbolic, and exponential solutions that are concurrently generated by the new auxiliary equation and the unified techniques.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!