The reproducibility of retention data and band profile characteristics was investigated for a series of columns packed with Luna C18 (2), a silica-based, reversed-phase adsorbent. High precision data were obtained and statistically compared among five columns from the same batch (column-to-column reproducibility) and nine columns from as many different batches (batch-to-batch reproducibility). These data were acquired under five different sets of chromatographic conditions, for a group of 30 neutral, acidic, and basic compounds selected as probes following an experimental protocol previously described. This work is part of a study on the precision of chromatographic analyses. Its purpose is to illustrate the contribution of the stationary phase reproducibility to this precision.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s0021-9673(99)01092-4 | DOI Listing |