Introduction: Lead systems that include an active pectoral pulse generator are now standard for initial defibrillator implantations. However, the optimal transvenous lead system and coil location for such active can configurations are unknown. The purpose of this study was to evaluate the benefit and optimal position of a superior vena cava (SVC) coil on defibrillation thresholds with an active left pectoral pulse generator and right ventricular coil.
Methods And Results: This prospective, randomized study was performed on 27 patients. Each subject was evaluated with three lead configurations, with the order of testing randomized. Biphasic shocks were delivered between the right ventricular coil and an active can alone (unipolar), or an active can in common with the proximal coil positioned either at the right atrial/SVC junction (low SVC) or in the left subclavian vein (high SVC). Stored energies at defibrillation threshold were higher for the single-coil, unipolar configuration (11.2 +/- 6.6 J) than for the high (8.9 +/- 4.2 J) or low (8.5 +/- 4.2 J) SVC configurations (P < 0.01). Moreover, 96% of subjects had low (< or = 15 J) thresholds with the SVC coil in either position compared with 81% for the single-coil configuration. Shock impedance (P < 0.001) was increased with the unipolar configuration, whereas peak current was reduced (P < 0.001).
Conclusion: The addition of a proximal transvenous coil to an active can unipolar lead configuration reduces defibrillation energy requirements. The position of this coil has no significant effect on defibrillation thresholds.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8167.2000.tb00731.x | DOI Listing |
Future Cardiol
January 2025
Department of Internal Medicine, Dow University of Health Sciences, Karachi, Pakistan.
Background: Implantable cardioverter-defibrillators (ICDs) are essential for reducing sudden cardiac death in patients at risk of ventricular arrhythmias. The choice of ICD lead - single-coil or dual-coil - can influence device performance and patient outcomes. This meta-analysis evaluates the comparative efficacy and safety of single-coil versus dual-coil ICD leads to inform clinical decision-making.
View Article and Find Full Text PDFPharmacoecon Open
January 2025
Optimax Access Ltd, Kenneth Dibben House, Enterprise Rd, Chilworth, Southampton University Science Park, Southampton, UK.
Background: Patients with a left ventricular ejection fraction ≤ 35% are at increased risk of sudden cardiac death (SCD) within the first months after a myocardial infarction (MI). The wearable cardioverter defibrillator (WCD) is an established, safe and effective solution which can protect patients from SCD during the first months after an MI, when the risk of SCD is at its peak. This study aimed to evaluate the cost-effectiveness of WCD combined with guideline-directed medical therapy (GDMT) compared to GDMT alone, after MI in the English National Health Service (NHS).
View Article and Find Full Text PDFStrahlenther Onkol
January 2025
TUM School of Medicine and Health, Department of Radiation Oncology, Technische Universität München (TUM), Klinikum rechts der Isar, Munich, Germany.
Purpose: Increasing life expectancy and advances in cancer treatment will lead to more patients needing both radiation therapy (RT) and cardiac implantable electronic devices (CIEDs). CIEDs, including pacemakers and defibrillators, are essential for managing cardiac arrhythmias and heart failure. Telemetric monitoring of CIEDs checks battery status, lead function, settings, and diagnostic data, thereby identifying software deviations or damage.
View Article and Find Full Text PDFJ Magn Reson Imaging
December 2024
Department of Radiology, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China.
Background: In arrhythmogenic cardiomyopathy (ACM), left ventricle-dominant presentation has poorer outcomes than right-dominant presentation, suggesting that interventricular functional disparity might play a role in patients' prognosis. However, the prognostic impact of ventricular functional discordance in ACM patients remains unknown.
Purpose: To assess whether ventricular functional disparity measured as ventricular discordance index, defined as the ratio of right-ventricular ejection fraction (RVEF) to left-ventricular ejection fraction (LVEF), might reveal prognostic disparities between phenotypes and offer added risk stratification value.
Pacing Clin Electrophysiol
January 2025
Department of Cardiology II - Electrophysiology, University Hospital Münster, Münster, Germany.
Background: Noninferiority of omitting intraoperative defibrillation threshold (DFT) testing has been documented for transvenous implantable cardioverter defibrillators (ICD) whereas data for the subcutaneous-ICD (S-ICD) regarding the need for DFT testing, especially during S-ICD generator replacement, is not available.
Methods: A total of 112 consecutive patients who underwent S-ICD generator replacement and routine testing were included in this retrospective single-center study and analyzed regarding the outcome of intraoperative DFT.
Results: The majority of patients (87.