Purpose: The authors, forming part of a multicenter project funded by the European Community, summarize the validation of a tissue-mimicking flow Doppler test object and of procedures for testing medical diagnostic Doppler equipment. The results of the project are expected to contribute to a future international IEC Standard concerning flow Doppler test objects (Draft IEC 61685 Standard) and for the European Medical Device Directive (MD 93/42/EEC).

Methods: Within this project a test protocol was developed that includes a set of different procedures, suitable for checking Spectral and Colour Doppler systems. The performance parameters for describing the image quality as well as the accuracy and the correct functioning of a system are in accordance with the definitions made in the Draft IEC 61685 Standard.

Results: A survey of the design and materials used for this Doppler test object will be presented with a special emphasis on the suitability of the procedures for routine measurement of performance parameters in hospitals.

Conclusion: The test object satisfies the requirements of the Draft IEC 61685 Standard. The test procedures in combination with this test object can be used for checking different transducer models with nominal frequencies between 2.5-10.0 MHz.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-1999-8918DOI Listing

Publication Analysis

Top Keywords

test object
20
draft iec
16
iec 61685
16
doppler test
12
test
8
flow doppler
8
61685 standard
8
performance parameters
8
doppler
6
iec
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!