[A multileaf collimator: additional quality assurance in a GE linear accelerator].

Strahlenther Onkol

Strahlentherapie und Radioonkologie, Klinikum Darmstadt.

Published: December 1999

Background: In comparison to a conventional collimator, a multileaf collimator demands a great deal of quality assurance procedures due to its large number of leaves. A concept for daily quality assurance is presented, mainly concerning the positioning accuracy of the leaves.

Material And Methods: Two leaf configurations including maximal opening as well as overtravel of single leaves, at a maximal opening of the jaws, are transmitted online in daily exchange from our record- and verify system to the linac. Aiming at a special test phantom a visual control of the positioning accuracy is performed. The leaf positioning is documented by an electronic portal imaging system and is compared with a reference shot by superposition of a grid.

Results: This method of quality assurance offers a fast and effective possibility to guarantee the proper function of the whole system by simulating the routine treatment situation.

Conclusions: Compared to a conventional collimator only a slightly greater workload is needed for quality assurance of a multileaf collimator.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s000660050050DOI Listing

Publication Analysis

Top Keywords

quality assurance
20
multileaf collimator
12
conventional collimator
8
positioning accuracy
8
maximal opening
8
quality
5
assurance
5
collimator additional
4
additional quality
4
assurance linear
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!