Background: A heparin/protamine titration system for measurement of heparin levels (Hepcon) is promising for efficient anticoagulation during cardiopulmonary bypass (CPB).

Methods: Fifty-seven patients subjected to CPB were divided into two groups, control (n = 24) and Hepcon groups (n = 33). The Hepcon group was further divided into three subgroups according to perfusion temperature. For the control group, conventional administration of an anticoagulant (300 IU/kg of heparin) and reversal protocol (heparin 1: protamine 1) was performed. For the Hepcon group, a heparin dose-response assay directed the initial dose of heparin. Hepcon also determined the dose of protamine by the titration. The initial dose of heparin in the control group (300 IU/kg) was statistically less than that of Hepcon group (360+/-80 IU/kg).

Results: In the Hepcon group, the sensitivity to heparin was correlated with coagulation time (r = -0.78) and antithrombin III levels (r = 0.70), and individual difference of sensitivity resulted in a wide range of dosage (160 to 490 IU/kg). A strong correlation was observed between plasma and whole blood concentration of heparin (r = 0.86). However, they did not correlate with ACT values. Perfusion temperature didn't affect the heparin level, but did the ACT value. In the Hepcon group, the dose of protamine was significantly less and adverse events were rare.

Conclusions: In conclusion, whole blood heparin measurements correlated well with plasma heparin concentration. Protamine titration of heparin reduced the dose of protamine and decreased the chance of adverse reactions.

Download full-text PDF

Source

Publication Analysis

Top Keywords

hepcon group
20
heparin
13
dose protamine
12
measurement heparin
8
heparin concentration
8
cardiopulmonary bypass
8
hepcon
8
perfusion temperature
8
control group
8
300 iu/kg
8

Similar Publications

Introduction: Individualized heparin and protamine management is increasingly used as a strategy to reduce coagulation activation and bleeding complications. While it is associated with increased heparin requirements during Cardiopulmonary Bypass (CPB), the impact on protamine administration remains controversial. We aim to investigate the effect of heparin level-guided monitoring on protamine dosing during cardiac surgery where low-anticoagulation protocols are implemented.

View Article and Find Full Text PDF

Introduction: Patients with Antiphospholipid Syndrome (APS) undergoing cardiopulmonary bypass (CPB) surgery are at increased risk for thrombotic and hemorrhagic complications. Anticoagulation during CPB is typically monitored with activated clotting time (ACT) which may be falsely prolonged in patients with APS. The Hepcon Hemostasis Management System quantitatively determines the whole blood heparin concentration through heparin/protamine titration.

View Article and Find Full Text PDF

Background: A multidisciplinary Quality Assurance/Performance Improvement study to identify the incidence of "heparin rebound" in our adult cardiac surgical population instead detected a thromboelastometry pattern suggestive of initial protamine overdose in 34% despite Hepcon-guided anticoagulation management. Analysis of our practice led to an intervention that made an additional lower-range Hepcon cartridge available to the perfusionists.

Methods: One year later, an IRB-approved retrospective study was conducted in >500 patients to analyze the effects of the intervention, specifically focusing on the impact of the initial protamine dose accuracy and 18-h mediastinal chest tube drainage (MCTd).

View Article and Find Full Text PDF

Purpose: High-dose heparin has been suggested to reduce consumption coagulopathy.

Materials And Methods: In a randomized, blinded, prospective trial of patients undergoing elective, complex cardiac surgery with cardiopulmonary bypass, patients were randomized to one of three groups: 1) high-dose heparin (HH) receiving an initial heparin dose of 450 u/kg, 2) heparin concentration monitoring (HC) with Hepcon Hemostasis Management System (HMS; Medtronic, Minneapolis, MN, USA) monitoring, or 3) a control group (C) receiving a standard heparin dose of 300 u/kg. Primary outcome measures were blood loss and transfusion requirements.

View Article and Find Full Text PDF

Purpose: We evaluated the efficacy of hemostatic therapy based on point-of-care (POC) testing in patients undergoing cardiac surgery.

Methods: This was a single-institution, prospective, randomized, double-blinded study. In step 1, 90 patients scheduled for elective cardiac surgery underwent complete blood count and fibrinogen measurements at baseline, after cardiopulmonary bypass (CPB) initiation (CPB start), just after CPB end, and in the intensive care unit (ICU).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!