Using a panel of monoclonal antibodies against human myoglobin (Mb), we have shown that the sensitivity of antigen-capture ELISA can be significantly increased by simultaneous immobilization of two cooperating capture monoclonal antibodies on a solid phase. This method ("triple-site ELISA") uses three monoclonal antibodies to different epitopes of the same antigen (two capture/one tracer) unlike the traditional double-site assay using one capture and one tracer monoclonal antibody. We developed double- and triple-site ELISA for Mb by varying the capture and tracer monoclonal antibodies. Triple-site assays showed 4-6-fold increase in sensitivity compared to the double-site assays. A model for this effect is suggested; according to the model, in triple-site ELISA, high-affinity cyclic configurations can be formed by an antigen, two capture monoclonal antibodies, and the surface of the solid phase.
Download full-text PDF |
Source |
---|
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!