Background: Dacron vascular prostheses are associated with thromboembolic complications and inflammatory responses; impregnation with bovine collagen reportedly stimulates additional inflammatory/immunologic complications. The Novacor (Baxter Healthcare Corp., Oakland, CA, USA) left ventricular assist system uses Dacron inflow and collagen-impregnated Dacron outflow prostheses.

Methods: Explanted inflow and outflow prostheses were evaluated for inflammatory/immunologic, hemostatic, anticoagulant, and fibrinolytic pathways. Non-implanted prostheses immersed in whole blood or plasma were used as controls.

Results: Immunoglobulins and complement components were observed in all prostheses with activated macrophages being present only in implanted prostheses. Antithrombin III was observed in all prostheses whereas fibrin, tissue plasminogen activator, and alpha-2 plasmin inhibitor were present only in implanted prostheses. Endothelial and smooth muscle cells associated with vascular structures containing collagen type IV and laminin were observed solely in implanted prostheses.

Conclusion: An inflammatory response occurs and key components of hemostatic, anticoagulant, and fibrinolytic pathways are present within implanted prostheses. These processes are accompanied by endothelial and smooth muscle cell infiltration which appear to lay the foundation for neovessel development.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s1053-2498(99)00003-0DOI Listing

Publication Analysis

Top Keywords

implanted prostheses
12
prostheses
9
vascular prostheses
8
left ventricular
8
ventricular assist
8
assist system
8
hemostatic anticoagulant
8
anticoagulant fibrinolytic
8
fibrinolytic pathways
8
observed prostheses
8

Similar Publications

Background: Dislocation remains a common complication following total hip arthroplasty (THA). Previous literature has shown that the femoral head-to-neck ratio is essential in hip motion, function, and stability. While large femoral heads and dual mobility bearings have been developed to improve stability, it remains unknown if the ratio between femoral head size to acetabular cup size also plays a role in stability.

View Article and Find Full Text PDF

Background: Coronal lower extremity malalignment and improper implant position have been described as risk factors for aseptic tibial loosening following primary total knee arthroplasty (TKA). However, several prior studies have shown no association between alignment and implant loosening. Meanwhile, there is increasing interest in kinematic alignment.

View Article and Find Full Text PDF

Impact of central corneal astigmatism on postoperative visual outcomes in patients undergoing trifocal intraocular lens implantation.

BMC Ophthalmol

January 2025

Department of Ophthalmology, Eye & ENT Hospital, Fudan University, 83 Fenyang Road, Xuhui District, Shanghai, 200031, China.

Background: To investigate the impact of central corneal astigmatism on postoperative visual outcomes in patients with trifocal intraocular lens (IOL) implantation.

Methods: This retrospective study included 278 eyes of 278 patients who underwent uneventful cataract surgery with implantation of the trifocal IOL (AT LISA tri 839MP). Patients were divided into two groups according to the total corneal refractive power (TCRP) in 3 mm zone centered on the corneal apex: low astigmatism group, TCRP ≤ 0.

View Article and Find Full Text PDF

Short-stem total hip arthroplasty (SHA) has become popular because it preserves femoral bone stock and enables the use of short femoral stems in revision total hip arthroplasty (THA). However, no study has evaluated whether a short stem in revision THA, replacing a standard stem, can provide adequate primary stability to facilitate osseous integration. In this biomechanical study, a metaphyseal anchoring SHA (Tri-Lock BPS) stem and a standard THA (Corail) stem were implanted into ten composite femurs and loaded dynamically from 300 to 1700 N with 1 Hz.

View Article and Find Full Text PDF

Background: Transcatheter closure of percutaneous paravalvular leak (PVL) is a technically challenging procedure, especially after surgical mechanical valve replacements (SMVR), as the risk of interference with the prosthetic valve discs and the complex interventional techniques required for mitral PVL closure. Our study was designed to review the results with transcatheter closure of PVL after SMVR.

Methods: From January 2018 through December 2023, a total of 64 patients with PVL after SMVR underwent transcatheter closure with the help of preoperative 3-dimensional printing model and simulator for image evaluation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!