Background: The analytical validation of multiple lots of the PRO-Trac II ELISA (DiaSorin) for the determination of tacrolimus in whole blood is described.

Methods: The analytical parameters assessed included analytical sensitivity, dilution linearity, functional sensitivity, values in samples containing no tacrolimus, intra- and interassay precision, supplementation and recovery, metabolite cross-reactivity, interference studies, and method comparisons HPLC-tandem mass spectrometry (HPLC/MS/MS) and the IMx Tacrolimus II multiparticle enzyme immunoassay. Where appropriate, assessments were performed according to NCCLS guidelines.

Results: The mean analytical detection limit was <0.25 microg/L for all lots, whereas the functional sensitivity was 1.0 microg/L. Excellent linear correlation (r = 0.985) was observed for dilution linearity. The intraassay imprecision was <7%, and the total imprecision by ANOVA was <10%. Recovery was 109% +/- 11%. Metabolite cross-reactivity was consistent with previous reports for this antibody. No interference was observed for 35 tested drugs. Method comparison with HPLC/MS/MS showed no statistically significant differences. Samples exhibited stability through four freeze/thaw cycles and for 1 week at room temperature.

Conclusion: These data demonstrate that the PRO-Trac II ELISA is a robust, accurate, and precise tool for the assessment and management of tacrolimus blood concentrations in transplant patients.

Download full-text PDF

Source

Publication Analysis

Top Keywords

analytical validation
8
pro-trac elisa
8
determination tacrolimus
8
analytical
5
validation pro-trac
4
elisa determination
4
tacrolimus
4
tacrolimus fk506
4
fk506 blood
4
blood background
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!