Purpose: The purpose of this study was to evaluate prospectively the results of the bifurcated Vanguard endovascular graft for abdominal aortic aneurysm (AAA) repair.

Methods: Seventy-five patients, with a median age of 69.6 years (range, 48 to 88 years) and asymptomatic AAAs, were recruited in 14 French vascular institutions. An independent committee validated the indications for endovascular repair, and all the implantations were supervised by a well-trained medico-technical assistant. Further independent committees reviewed patient data, clinical data, and imaging follow-up examination. The main endpoints were implantation success, mortality, morbidity, reinterventions, and aneurysm evolution assessed with serial computed tomographic (CT) scanning.

Results: All the grafts were successfully implanted, resulting in a 100% success rate on an intent-to-treat basis. At discharge, there were no deaths, six significant local complications (8%) that necessitated surgery, no vascular complications, and six systemic complications (8%). The average durations of intensive care unit and hospital stays were 26 +/- 6 hours and 6 +/- 2.54 days, respectively. Predischarge CT scan results showed five type I and 18 type II endoleaks (total, 30%). At the end of the follow-up period (mean duration, 18.35 +/- 4.12 months; range, 17 days to 24 months), seven patients (9%) had died: one from sepsis, five from unrelated causes, and one from aneurysm rupture. The 2-year cumulative survival rate was 86% +/- 5.9%. Twenty-one subsequent endovascular or vascular procedures were necessitated (28%) in 17 patients (23%) to treat graft limb occlusion or stenosis (n = 9 patients) or to seal an endoleak (n = 8 patients). The 2-year cumulative survival rate free of reintervention was 67% +/- 7%. On CT scans, the mean AAA diameter decreased from 54 mm +/- 8.9 (range, 45 to 80 mm) before surgery to 51.6 mm +/- 9.1 at 6 months and to 43.4 mm +/- 4.4 at the end of the follow-up period (P =.001). Persistent endoleak was significantly associated with an increase in diameter (4 of 5 [80%] vs 1 of 47 [2%]; P =.001).

Conclusion: In selected patients, the bifurcated Vanguard endovascular graft may be implanted with a low mortality and morbidity rate and a favorable mid-term survival rate. The decrease of the aneurysm size is a strong argument in favor of the efficiency of the device. However, lasting endoleaks with increased aneurysm diameter and occurrence of limb graft stenosis or occlusion raise concerns and justify a careful long-term follow-up monitoring of all patients who undergo treatment with endovascular technique.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0741-5214(99)70130-8DOI Listing

Publication Analysis

Top Keywords

endovascular graft
12
survival rate
12
graft abdominal
8
abdominal aortic
8
aortic aneurysm
8
bifurcated vanguard
8
vanguard endovascular
8
mortality morbidity
8
+/-
8
follow-up period
8

Similar Publications

Subarachnoid hemorrhage (SAH) is a type of hemorrhagic stroke with high morbidity, mortality and disability, and early brain injury (EBI) after SAH is crucial for prognosis. Recently, stem cell therapy has garnered significant attention in the treatment of neurological diseases. Compared to other stem cells, dental pulp stem cells (DPSCs) possess several advantages, including abundant sources, absence of ethical concerns, non-invasive procurement, non-tumorigenic history and neuroprotective potential.

View Article and Find Full Text PDF

Thoracoabdominal aortic aneurysm (TAAA) repair remains one of the most challenging procedures and is associated with high mortality and complication rates. Careful consideration of the surgical strategy is essential, particularly in cases involving extensive replacement and high-risk patients. A 61-year-old man with a 55-mm TAAA was referred for surgical treatment.

View Article and Find Full Text PDF

Objective: The aim of this study was to perform a comparative pre-clinical evaluation of a new fenestrated endovascular aneurysm repair dedicated stent using perfused 3D printed patient anatomies.

Methods: The test setup included a pulsatile pump set to reproduce human haemodynamics, four 3D printed pararenal aneurysms connected to a bench test, and four corresponding fenestrated grafts (Zenith Fenestrated; Cook, Bloomington, IN, USA). Bridging stents were sized based on analysis of patients computed tomography scans and included either four standard covered stents (BeGraft [BG]; Bentley, Hechingen, Germany) or four similar covered stent grafts mounted on a modified balloon to allow stent implantation and flaring in one step (BeFlared [BF]; Bentley).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!