Purpose: To assess the ability of an extravasation detection accessory (EDA) to detect clinically important (> or = 10 mL) extravascular injection of iodinated contrast material delivered with an automated power injector.
Materials And Methods: Five hundred patients referred for contrast material-enhanced body computed tomography (CT) participated in a prospective, multi-institutional, observational study in which the EDA was used to identify and interrupt any injection associated with clinically important extravasation. The presence or absence of extravasation was definitively established with helical CT at the injection site (injection rate, from 1.0 to 5.0 mL/sec; mean, 2.9 mL/sec; median, 3.0 mL/sec).
Results: There were four true-positive (extravasation volumes, 13-18 mL), 484 true-negative, 12 false-positive, and no false-negative cases. The prevalence of overall and clinically important (> or = 10 mL) extravasation was 3.6% (18 of 500 cases) and 0.8% (four of 500 cases), respectively. The EDA had a sensitivity of 100% (four of four cases; 95% CI: 51%, 100%) and a specificity of 98% (484 of 496 cases; 95% CI: 96%, 99%) in the detection of clinically important extravasation.
Conclusion: The EDA is easy to use, safe, and accurate in the monitoring of intravenous injections for extravasation, which could prove especially useful in high-flow-rate CT applications.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1148/radiology.212.2.r99au14431 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!