Study Design: A mechanical testing standard for anterior thoracolumbar instrumentation systems was introduced, using a synthetic model. Twelve recent instrumentation systems were tested in static and fatigue modes.
Objectives: To establish the testing standard for anterior thoracolumbar instrumentation systems using a synthetic model and to evaluate the static and fatigue biomechanical properties of 12 anterior thoracolumbar instrumentation systems.
Summary Of Background Data: Although numerous studies have been performed to evaluate the biomechanics of anterior spinal instrumentation using a cadaveric or animal tissue, problems of specimen variation, lack of reproducibility, and inability to perform fatigue testing have been pointed out. In no studies has a precise synthetic testing standard for anterior thoracolumbar instrumentation systems been described.
Methods: An ultra-high-molecular-weight polyethylene cylinder was designed according to the anatomic dimensions of the vertebral body. Two cylinders spanned by spinal instrumentation simulated a total corpectomy defect, and a compressive lateral bending load was applied. The instrumentation assembly was precisely standardized. The static destructive and fatigue tests up to 2 million cycles at three load levels were conducted, followed by the failure mode analysis. Twelve anterior instrumentation systems, consisting of five plate and seven rod systems were compared in stiffness, bending strength, and cycles to failure.
Results: Static and fatigue test parameters both demonstrated highly significant differences between devices. The stiffness ranged from 280.5 kN/m in the Synthes plate (Synthes, Paoli, PA) to 67.9 kN/m in the Z-plate ATL (SofamorDanek, Memphis, TN). The Synthes plate and Kaneda SR titanium (AcroMed, Cleveland, OH) formed the highest subset in bending strength of 1516.1 N and 1209.9 N, respectively, whereas the Z-plate showed the lowest value of 407.3 N. There were no substantial differences between plate and rod devices. In fatigue, only three systems: Synthes plate, Kaneda SR titanium, and Olerud plate (Nord Opedic AB, Sweden) withstood 2 million cycles at 600 N. The failure mode analysis demonstrated plate or bolt fractures in plate systems and rod fractures in rod systems.
Conclusions: The biomechanical testing standard for anterior thoracolumbar instrumentation systems was successfully designed. It provided a repeatable and consistent experimental condition and controlling dimensional and surgical factors. The comparison of 12 instrumentation systems highlights the importance of mechanically balanced device design without a weak link in the development of instrumentation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/00007632-199907150-00004 | DOI Listing |
Water Sci Technol
January 2025
Qingdao Branch of Luoyang Ship Material Research Institute, 149-1, Zhuzhou Road, Laoshan District, Qingdao, Shandong, China; Sunrui Marine Environment Engineering Co., Ltd, Qingdao, Shandong, China.
Nowadays, performance studies on the amperometric total residual oxidant (TRO) sensor are only in the bench test stage and have not been conducted under specific maritime conditions with Ballast Water Management System (BWMS). In this study, the application of the amperometric TRO sensor in land-based biological efficacy (BE) testing, operation and maintenance (O&M) testing, as well as shipboard (SB) testing, was explored by comparing with the existing di-phenylene-diamine (DPD) TRO sensor. The results showed that the average TRO measurement deviation between the amperometric sensor and the DPD sensor was within ±10% in valid BE test cycles and the O&M testing exceeding 47 operating hours.
View Article and Find Full Text PDFWater Sci Technol
January 2025
China Construction Fifth Engineering Division Co., Ltd, Changsha, Hunan 410004, China.
Road runoff underwent treatment using a filter filled with sludge from drinking water treatment plants to assess its capacity for removing dissolved organic matter (DOM). This evaluation utilized resin fractionation, gel permeation chromatography, three-dimensional excitation-emission matrix fluorescence spectroscopy, and UV-Visible spectroscopy. The filter demonstrated enhanced efficiency in removing dissolved organic carbon, achieving removal rates between 70 and 80%.
View Article and Find Full Text PDFLab Chip
January 2025
CAS Key Laboratory of Mechanical Behavior and Design of Materials, Department of Precision Machinery and Precision Instrumentation, University of Science and Technology of China, Hefei, 230027, P. R. China.
We propose a novel contactless droplet manipulation strategy that combines electrostatic tweezers (ESTs) with lubricated slippery surfaces. Electrostatic induction causes the droplet to experience an electrostatic force, allowing it to move with the horizontal shift of the EST. Because both the EST and the slippery operating platform prepared by a femtosecond laser exhibit a strong binding effect on droplets, the EST droplet manipulation features significant flexibility, high precision, and can work under various operating conditions.
View Article and Find Full Text PDFAim: The aim of the study is to compare the remaining dentinal thickness in the single-rooted tooth at the coronal, middle, and apical third using three different rotary instrumentation techniques using cone-beam computed tomography (CBCT).
Materials And Methodology: Sixty-eight noncarious single-rooted teeth were collected and decoronated at the level of cementoenamel junction with a diamond disc. All specimens were randomly divided into three experimental groups - Manual ProTaper Universal (PTU; Dentsply Maillefer, Ballaigues, Switzerland), Protaper Gold (PTG; Dentsply Maillefer), Waveone Gold (Dentsply Maillefer) and one control group of 17 teeth each.
JMIR Mhealth Uhealth
January 2025
Department of Learning and Workforce Development, The Netherlands Organisation for Applied Scientific Research, Soesterberg, Netherlands.
Background: Wearable sensor technologies, often referred to as "wearables," have seen a rapid rise in consumer interest in recent years. Initially often seen as "activity trackers," wearables have gradually expanded to also estimate sleep, stress, and physiological recovery. In occupational settings, there is a growing interest in applying this technology to promote health and well-being, especially in professions with highly demanding working conditions such as first responders.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!