[Is corneal thickness measurement reliable and useful?].

J Fr Ophtalmol

Service d'Ophtalmologie, CHU Dijon.

Published: March 1999

Purpose: We evaluated the accuracy and the reproducibility of central corneal thickness measurements using an ultrasound pachymeter, and its usefulness in clinical practice.

Methods: We calculated the intra-observer, inter-observer, and inter-session variability in control subjects (n = 38). We observed the diurnal variation (n = 8) and the role of surgery (n = 12) on central corneal thickness. We measured and compared central corneal thickness and intraocular pressure in 6 groups (control subjects n = 134, primary open-angle glaucoma n = 111, ocular hypertension n = 66, normal tension glaucoma n = 12, diabetes mellitus n = 62, corneal graft n = 27). We studied the influence of dorzolamide on central corneal thickness (n = 16).

Results: The intra-observer variability was 9 +/- 4 microns, whereas the inter-observer and inter-session variabilities were 4 microns and 5% m respectively. The central corneal thickness in patients with ocular hypertension (587 +/- 41 microns) was significantly greater than control subjects (548 +/- 32 microns) and all the other groups (p < 0.001). No influence of dorzolamide was observed on central corneal thickness.

Conclusion: Central corneal thickness assessed with an ultrasound pachymeter may be a useful and accurate method in selected patients whose intraocular pressure measurement does not correlate with other clinical findings.

Download full-text PDF

Source

Publication Analysis

Top Keywords

corneal thickness
28
central corneal
28
control subjects
12
+/- microns
12
corneal
8
ultrasound pachymeter
8
inter-observer inter-session
8
intraocular pressure
8
ocular hypertension
8
influence dorzolamide
8

Similar Publications

Purpose: To assess the repeatability of lipid layer thickness (LLT) measurement using the LipiView® interferometer after daily disposable contact lens (CL) wear and correlation with ocular comfort in soft contact lens wearers.

Methods: A prospective study was conducted over two consecutive months, wherein CL wearers (n = 20) wore either Somofilcon A or Verofilcon A daily disposable CLs in a crossover design, switching lenses after 1 month. The pre-corneal tear film LLT was measured at the end of each month after CLs had been worn for at least 6 h.

View Article and Find Full Text PDF

Purpose: This report presents a case of fungal keratitis treated with penetrating keratoplasty using a cryopreserved cornea, highlighting the successful maintenance of corneal transparency post infection resolution.

Observations: A 57-year-old man complaining of pain in the right eye was referred to our hospital. Although diagnosed with fungal keratitis, his corneal scraping indicated the presence of , and he was unresponsive to voriconazole, micafungin, and pimaricin treatments.

View Article and Find Full Text PDF

Keratoconus (KC) is a progressive corneal disorder resulting in severe visual impairment. We aimed to determine the prevalence and corneal tomographic characteristics of KC and keratoconus suspect (KCS) in a population-based study, and to construct discrimination models with or without corneal tomography. A total of 1,544 eyes (822 participants aged ≥35 years) were evaluated using data from the Yamagata Study (2015-2017).

View Article and Find Full Text PDF

Corneal blindness, a leading cause of visual impairment globally, has created a pressing need for alternatives to corneal transplantation due to the severe shortage of donor tissues. In this study, we present a novel interpenetrating network hydrogel composed of gelatin methacryloyl (GelMA) and oxidized carboxymethyl cellulose (OxiCMC) for bioprinting a biomimetic corneal stroma equivalent. We tested different combinations of GelMA and OxiCMC to optimize printability and subsequently evaluated these combinations using rheological studies for gelation and other physical, chemical, and biological properties.

View Article and Find Full Text PDF

Refractive surgery, which includes techniques such as Laser-Assisted In Situ Keratomileusis (LASIK), Photorefractive Keratectomy (PRK) and Small Incision Lenticule Extraction (SMILE), has revolutionized ophthalmology by offering advanced solutions for vision correction. However, the choice of the technique to be used in the individual patient is highly dependent on a thorough preoperative evaluation. This retrospective study aims to investigate how preoperative parameters, including corneal thickness, topography, and refraction, affect long-term post-operative clinical outcomes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!