Background: To decrease the complications associated with cardiopulmonary bypass, novel biomaterials have been introduced that may be less thrombogenic than standard synthetic surfaces.

Methods: Thirty-four patients undergoing coronary artery bypass grafting were randomized to bypass using either a control circuit or a circuit prepared "tip-to-tip" with a triblock-copolymer (polycaprolactone-polydimethylsiloxane-polycaprolactone).

Results: There was a progressive increase in thrombin generation in the control group during bypass, which was not seen in the test group. The test surface decreased the release of tissue plasminogen activator and plasmin-alpha2-antiplasmin complex formation (p<0.005). There was also an increased platelet count and a decreased platelet activation in the test group, as detected by GMP-140 expression and beta-thromboglobulin release (p = 0.017). There was also significantly more debris that accumulated on the arterial filter in the control group, as confirmed by scanning electron microscopy.

Conclusions: This clinical trial has demonstrated a significant difference in the hematologic effects of the test circuits, with evidence of platelet preservation, decreased fibrinolysis, and decreased thrombin generation. A larger trial would be necessary to establish the clinical relevance of these differences.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0003-4975(99)00068-5DOI Listing

Publication Analysis

Top Keywords

cardiopulmonary bypass
8
bypass
5
hematologic evaluation
4
evaluation cardiopulmonary
4
bypass circuits
4
circuits prepared
4
prepared novel
4
novel block
4
block copolymer
4
copolymer background
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!