Percutaneous fibrin sleeve stripping of failing haemodialysis catheters.

Nephrol Dial Transplant

Department of Radiology, West Glasgow Hospitals University NHS Trust, Gartnavel General Hospital, UK.

Published: March 1999

Background: One of the most frequent reasons for failure of haemodialysis lines is catheter blockage caused by fibrin sheath formation. We report our experience of percutaneous fibrin sheath stripping in treating this problem.

Methods: A consecutive series of failing haemodialysis catheters underwent percutaneous fibrin sheath stripping in an attempt to retrieve and prolong the life of the catheter. Immediate technical success, clinical success, and primary and secondary patency were measured based on clinical follow-up.

Results: Sixteen non-functional permanent haemodialysis lines in 15 patients underwent percutaneous fibrin sheath stripping on 21 occasions. Technical success rate was 100%. Catheter flow rates sufficient for initial dialysis were achieved in 12 (75%) lines. Successful percutaneous fibrin sheath stripping produced a mean catheter patency of 126 days (range 6-299 days).

Conclusions: Percutaneous fibrin sheath stripping is a simple, repeatable procedure that can usefully extend the life of a failing dialysis line. However, a randomized trial will be needed to evaluate its role compared with catheter replacement.

Download full-text PDF

Source
http://dx.doi.org/10.1093/ndt/14.3.688DOI Listing

Publication Analysis

Top Keywords

percutaneous fibrin
24
fibrin sheath
24
sheath stripping
20
failing haemodialysis
8
haemodialysis catheters
8
haemodialysis lines
8
underwent percutaneous
8
technical success
8
percutaneous
6
stripping
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!