The biocompatibility and degradation of self-reinforced poly-L-lactide (SR-PLLA) and polyglycolide (SR-PGA) miniscrews, vs titanium miniscrews, was studied in frontal bone osteotomies in 20 lambs, where they were used for plate fixation. At follow-up at 4, 6, 12, 26, 52 and 104 weeks, no clinical foreign body reaction, infection or other complications had occurred. Histologically, PGA material was hydrolyzed and fragmented at 4-6 weeks and was resorbed by 12 weeks, whereas the SR-PLLA miniscrews retained their integrity and holding power for 26 weeks and were mostly resorbed at 2 years. According to histological and histomorphometric analyses and plain film radiography, the degradation of PGA miniscrews was accompanied by a typical non-specific foreign-body reaction and initial transient osteolysis with decreased osteoid formation around the screw channel, but compensatory intense osteoid formation and bone remodelling followed after resorption of the polymer. The foreign body reactions to PLLA and titanium were considerably milder. All miniscrews were commendably strong and could be satisfactorily tightened against the plate. SR-PLLA miniscrews offer fixation stability for half a year, whereas rapidly degrading SR-PGA miniscrews may be used when short-term fixation is needed.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s1010-5182(99)80009-2 | DOI Listing |
J Neurosurg
May 1999
Department of Surgery, University of Helsinki, Finland.
Object: To determine the biocompatibility and suitability of resorbable plates and miniscrews, consolidation of symmetrical, bilateral frontal bone craniotomies that had been closed using various methods was studied in 20 growing lambs.
Methods: Bone fixation with a flexible, punched polylactide plate and four slowly degradable, self-reinforced poly-levolactide (SR-PLLA) or rapidly degradable, self-reinforced polyglycolide (SR-PGA) miniscrews (10 animals in each group) was compared intraindividually with rigid fixation by using a titanium miniplate and four miniscrews. Plain x-ray films, magnetic resonance images, histological studies, and histomorphometric studies were obtained at 4 to 104 weeks.
J Craniomaxillofac Surg
February 1999
Department of Surgery, University of Helsinki, Finland.
The biocompatibility and degradation of self-reinforced poly-L-lactide (SR-PLLA) and polyglycolide (SR-PGA) miniscrews, vs titanium miniscrews, was studied in frontal bone osteotomies in 20 lambs, where they were used for plate fixation. At follow-up at 4, 6, 12, 26, 52 and 104 weeks, no clinical foreign body reaction, infection or other complications had occurred. Histologically, PGA material was hydrolyzed and fragmented at 4-6 weeks and was resorbed by 12 weeks, whereas the SR-PLLA miniscrews retained their integrity and holding power for 26 weeks and were mostly resorbed at 2 years.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!