Accuracy and precision of a robotic sample processor.

J Int Fed Clin Chem

Armstrong Laboratory, Human Systems Center, Brooks Air Force Base, Texas 78235.

Published: September 1992

The accuracy and precision of the Packard Probe 1000, a computer-driven robotic sample processor, were determined using NCCLS Guideline 18-P, employing a rigorous gravimetric procedure. For volumes of 25, 100, and 200 muL, inaccuracy ranged from -0.5 to +0.6% (saline) and -0.8 to +0.7% (human serum), and imprecision from 0.1 to 0.6% (saline) and 0.02 to 0.8% (human serum), in the dip and sip mode of pipetting. For the same volumes of saline in the dispense-through pipetting mode, inaccuracy ranged from -0.1 to -0.9% and imprecision from 0.1 to 0.5%. To obtain this level of performance, the instrument's operating parameters were changed from the factory default settings. This automated liquid-handling system is capable of impressive accuracy and precision, but it must be calibrated in the same manner as any other volumetric device. Performance varies with the volume, type of liquid, and mode of pipetting. Inaccuracy of less than or equal to 1.0% and imprecision of less than =1.0% can be attained along with rapid, walkaway pipetting of calibrators, controls, and large numbers of patient samples.

Download full-text PDF

Source

Publication Analysis

Top Keywords

accuracy precision
12
robotic sample
8
sample processor
8
inaccuracy ranged
8
human serum
8
mode pipetting
8
precision robotic
4
processor accuracy
4
precision packard
4
packard probe
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!