We developed a sensitive and specific semi-automated liquid chromatography-electrospray mass spectrometric (HPLC-ESI-MS) assay for the simultaneous quantification of sirolimus and ciclosporin in blood. Following a simple protein precipitation step, the supernatants were injected into the HPLC system and extracted on-line. After column switching, the analytes were backflushed from the extraction column onto the analytical narrow-bore column and eluted into the ESI-MS system. The assay was linear from 0.4 to 100 microg/l sirolimus and from 2 to 1500 microg/l ciclosporin. The mean recoveries of sirolimus and ciclosporin were 98 and 96%, respectively. The mean interday precision/accuracy was 8.6%/-4.8% for sirolimus and 9.3%/-2.9% for ciclosporin.