Three electrosurgical tissue-sealing devices (EnSeal ETSDRC-01, LigaSure LS1500 and Thunderbeat TB-0535PC) were compared regarding sealing time (ST), maximum working temperature (WTmax) and the total (MTZtotal) as well as the collateral microscopic thermal injury zone (MTZcollat) using laparoscopic handpieces 5 mm in diameter on four types of tissue (liver, mesentery, cross striated muscle and spleen) in an in vivo porcine model. LigaSure had the lowest mean ST in spleen, mesentery, muscle and liver, followed by Thunderbeat and EnSeal with significant differences between all types of tissues and devices. The significantly lowest mean WTmax was obtained for EnSeal in mesentery, muscle and liver.
View Article and Find Full Text PDF