Purpose: This work aims to simulate clustered DNA damage from ionizing radiation and estimate the relative biological effectiveness (RBE) for radionuclide (rBT)- and electronic (eBT)-based surface brachytherapy through a hybrid Monte Carlo (MC) approach, using realistic models of the sources and applicators.
Methods: Damage from ionizing radiation has been studied using the Monte Carlo Damage Simulation algorithm using as input the primary electron fluence simulated using a state-of-the-art MC code, PENELOPE-2018. Two Ir rBT applicators, Valencia and Leipzig, one Co source with a Freiburg Flap applicator (reference source), and two eBT systems, Esteya and INTRABEAM, have been included in this study implementing full realizations of their geometries as disclosed by the manufacturer.
Purpose: In image-guided adaptive brachytherapy (IGABT) a quantitative evaluation of the dosimetric changes between fractions due to anatomical variations, can be implemented via rigid registration of images from subsequent fractions based on the applicator as a reference structure. With available treatment planning systems (TPS), this is a manual and time-consuming process. The aim of this retrospective study was to automate this process.
View Article and Find Full Text PDFPurpose: Recently, we introduced a bi-objective optimization approach based on dose-volume indices to automatically create clinically good HDR prostate brachytherapy plans. To calculate dose-volume indices, a reconstruction algorithm is used to determine the 3D organ shape from 2D contours, inevitably containing settings that influence the result. We augment the optimization approach to quickly find plans that are robust to differences in 3D reconstruction.
View Article and Find Full Text PDFPurpose: Bi-objective simultaneous optimization of catheter positions and dwell times for high-dose-rate (HDR) prostate brachytherapy, based directly on dose-volume indices, has shown promising results. However, optimization with the state-of-the-art evolutionary algorithm MO-RV-GOMEA so far required several hours of runtime, and resulting catheter positions were not always clinically feasible. The aim of this study is to extend the optimization model and apply GPU parallelization to achieve clinically acceptable computation times.
View Article and Find Full Text PDFThe surface brachytherapy Task Group report number 253 discusses the common treatment modalities and applicators typically used to treat lesions on the body surface. Details of commissioning and calibration of the applicators and systems are discussed and examples are given for a risk-based analysis approach to the quality assurance measures that are necessary to consider when establishing a surface brachytherapy program.
View Article and Find Full Text PDFThree different correction factors for measurements with the parallel-plate ionization chamber PTW T34013 on the Esteya electronic brachytherapy unit have been investigated. This chamber type is recommended by AAPM TG-253 for depth-dose measurements in the 69.5 kV x-ray beam generated by the Esteya unit.
View Article and Find Full Text PDFPurpose: To evaluate the use of the absorbed depth-dose as a surrogate of the half-value layer in the calibration of a high-dose-rate electronic brachytherapy (eBT) equipment. The effect of the manufacturing tolerances and the absorbed depth-dose measurement uncertainties in the calibration process are also addressed.
Methods: The eBT system Esteya (Elekta Brachytherapy, Veenendaal, The Netherlands) has been chosen as a proof-of-concept to illustrate the feasibility of the proposed method, using its 10 mm diameter applicator.
Purpose: The purpose of this study is to improve upon a recently introduced bi-objective treatment planning method for prostate high-dose-rate (HDR) brachytherapy (BT), both in terms of resulting plan quality and runtime requirements, to the extent that its execution time is clinically acceptable.
Methods: Bi-objective treatment planning is done using a state-of-the-art multiobjective evolutionary algorithm, which produces a large number of potential treatment plans with different trade-offs between coverage of the target volumes and sparing organs at risk. A graphics processing unit (GPU) is used for large-scale parallelization of dose calculations and the calculation of the dose-volume (DV) indices of potential treatment plans.
Prostate brachytherapy treatment using permanent implantation of low-energy (LE) low-dose rate (LDR) sources is successfully and widely applied in Europe. In addition, seeds are used in other tumour sites, such as ophthalmic tumours, implanted temporarily. The calibration issues for LE-LDR photon emitting sources are specific and different from other sources used in brachytherapy.
View Article and Find Full Text PDFPurpose: To investigate the variation in computed dose-volume (DV) indices for high-dose-rate (HDR) prostate brachytherapy that can result from typical differences in computation settings in treatment planning systems (TPSs).
Methods: Five factors were taken into account: number of dose-calculation points, radioactive source description, interpolation between delineated contours, intersections between delineated organ contours, and organ shape at the top and bottom contour using either full or partial slice thickness. Using in-house developed software, the DV indices of the treatment plans of 26 patients were calculated with different settings, and compared to a baseline setting that closely followed the default settings of the TPS used in our medical center.
Purpose: The purpose of this work is threefold: First, to obtain the phase space of an electronic brachytherapy (eBT) system designed for surface skin treatments. Second, to explore the use of some efficiency enhancing (EFEN) strategies in the determination of the phase space. Third, to use the phase space previously obtained to perform a dosimetric characterization of the Esteya eBT system.
View Article and Find Full Text PDFPurpose: The direction modulated brachytherapy (DMBT) magnetic resonance-compatible tandem applicator, made from a tungsten alloy rod, has six symmetric peripheral grooves, designed specifically to enhance intensity modulation capacity through achieving directional radiation dose profiles. In this work, the directional dose distributions of the DMBT tandem were modeled and calculated with the Oncentra Brachy advanced collapsed cone engine (ACE), which was validated against Monte Carlo (MC) calculations.
Methods And Material: The prototype 3D tandem applicator model was created for use in the Oncentra Brachy treatment planning system.
Purpose: To investigate the dose calculation accuracy of the Advanced Collapsed cone Engine (ACE) algorithm for ocular brachytherapy using a COMS plaque loaded with I-125 seeds for two heterogeneous patient tissue scenarios.
Methods: The Oncura model 6711 I-125 seed and 16 mm COMS plaque were added to a research version (v4.6) of the Oncentra Brachy (OcB) treatment planning system (TPS) for dose calculations using ACE.
Purpose: To investigate the dose calculation accuracy in water medium of the Advanced Collapsed cone Engine (ACE) for three sizes of COMS eye plaques loaded with low-energy I-125 seeds.
Methods: A model of the Oncura 6711 I-125 seed was created for use with ACE in Oncentra Brachy (OcB) using primary-scatter separated (PSS) point dose kernel and Task Group (TG) 43 datasets. COMS eye plaque models of diameters 12, 16, and 20 mm were introduced into the OcB applicator library based on 3D CAD drawings of the plaques and Silastic inserts.
A new electronic brachytherapy unit from Elekta, called Esteya(®), has recently been introduced to the market. As a part of the standards in radiation oncology, an acceptance testing and commissioning must be performed prior to treatment of the first patient. In addition, a quality assurance program should be implemented.
View Article and Find Full Text PDFBackground And Purpose: A substantial reduction of uncertainties in clinical brachytherapy should result in improved outcome in terms of increased local control and reduced side effects. Types of uncertainties have to be identified, grouped, and quantified.
Methods: A detailed literature review was performed to identify uncertainty components and their relative importance to the combined overall uncertainty.