Background: To reduce myocardial damage caused by implantable cardioverter defibrillator (ICD) shock, the left axilla was studied as an alternative pulse generator implantation site, and compared with the traditional implantation site, the left anterior chest.
Methods and results: Computer simulation was used to study the defibrillation conduction pattern and estimate the simulated defibrillation threshold (DFT) and myocardial damage when pulse generators were placed in the left axilla and left anterior chest, respectively; pulse generators were also newly implanted in the left axilla (n=30) and anterior chest (n=40) to compare the corresponding DFT. On simulation, when ICD generators were implanted in the left axilla, compared with the left anterior chest, the whole heart may be defibrillated with a lower defibrillation energy (left axilla 6.
Background: The need for ways to minimize the number of implantable cardioverter-defibrillator (ICD) shocks is increasing owing to the risk of its adverse effects on life expectancy. Studies have shown that a longer detection time for ventricular tachyarrhythmia reduces the safety of therapies, in terms of syncope and mortality, but not substantially in terms of the success rate. We aimed to evaluate the effects of increased number of intervals to detect (NID) VF on the safety of ICD shock therapy and on the reduction of inappropriate shocks.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2015
Background: Pacemaker generators are routinely implanted in the anterior chest. However, where to place the generator may need to be considered from the mental, functional, and cosmetic standpoints.
Methods: In this study, we performed the left axillary pacemaker generator implantation with a direct puncture of the left axillary vein in 40 consecutive patients, and evaluated the late safety and efficacy of this implantation.