Photon-counting detector CT (PCD-CT) is a new technology that has multiple diagnostic benefits including increased spatial resolution, iodine signal, and radiation dose efficiency, as well as multi-energy imaging capability, but which also has unique challenges in abdominal imaging. The purpose of this work is to summarize key features, technical parameters, and terms, which are common amongst current abdominopelvic PCD-CT systems and to propose standardized terminology (where none exists). In addition, user-selectable protocol parameters are highlighted to facilitate both scientific evaluation and early clinical adoption.
View Article and Find Full Text PDFBackground: Direct conversion x-ray Photon Counting Detectors (PCD) are posed to play a vital role in future medical imaging devices such as Computed Tomography (CT) scanners. PCD are expected to improve current CT technology on several fronts, such as resolution, dose utilization, and spectral performance. However, they are not readily expected to improve the handling of object scatter, one of the major sources of image artifacts in CT technology.
View Article and Find Full Text PDFPurpose: The purpose of this study was to investigate the feasibility of gadolinium-K-edge-angiography (angio-Gd-K-edge) with gadolinium-based contrast agents (GBCAs) as obtained with spectral photon counting CT (SPCCT) in atherosclerotic rabbits.
Materials And Methods: Seven atherosclerotic rabbits underwent angio-SPCCT acquisitions with two GBCAs, with similar intravenous injection protocol. Conventional and angio-Gd-K-edge images were reconstructed with the same parameters.
Objectives: To evaluate the quality of virtual monochromatic images (VMIs) from spectral photon-counting CT (SPCCT) and two energy-integrating detector dual-energy CT (EID-DECT) scanners from the same manufacturer, for the coronary lumen.
Methods: A 21-cm section of the Mercury v4.0 phantom was scanned using a cardiac CT protocol.
Spectral photon-counting computed tomography (SPCCT) technology holds great promise for becoming the next generation of computed tomography (CT) systems. Its technical characteristics have many advantages over conventional CT imaging. For example, SPCCT provides better spatial resolution, greater dose efficiency for ultra-low-dose and low-dose protocols, and tissue contrast superior to that of conventional CT.
View Article and Find Full Text PDFRationale And Objectives: A method for visualizing and analyzing the complete information contained in spectral CT scans using two-dimensional histograms (i.e. Material Attenuation Decomposition plots - MADplots) of the water-photoelectric attenuation versus water-scatter attenuation at the cohort (combination of multiple studies across patients), examination, series, slice, and organ/ROI levels is described.
View Article and Find Full Text PDFBackground Spatial resolution, soft-tissue contrast, and dose-efficient capabilities of photon-counting CT (PCCT) potentially allow a better quality and diagnostic confidence of coronary CT angiography (CCTA) in comparison to conventional CT. Purpose To compare the quality of CCTA scans obtained with a clinical prototype PCCT system and an energy-integrating detector (EID) dual-layer CT (DLCT) system. Materials and Methods In this prospective board-approved study with informed consent, participants with coronary artery disease underwent retrospective electrocardiographically gated CCTA with both systems after injection of 65-75 mL of 400 mg/mL iodinated contrast agent at 5 mL/sec.
View Article and Find Full Text PDFAims: To evaluate spectral photon-counting CT's (SPCCT) objective image quality characteristics in vitro, compared with standard-of-care energy-integrating-detector (EID) CT.
Methods: We scanned a thorax phantom with a coronary artery module at 10 mGy on a prototype SPCCT and a clinical dual-layer EID-CT under various conditions of simulated patient size (small, medium, and large). We used filtered back-projection with a soft-tissue kernel.
Objectives: The aim of this study is to compare the image quality of in vivo coronary stents between an energy integrating detectors dual-layer computed tomography (EID-DLCT) and a clinical prototype of spectral photon counting computed tomography (SPCCT).
Materials And Methods: In January to June 2021, consecutive patients with coronary stents were prospectively enrolled to undergo a coronary computed tomography (CT) with an EID-DLCT (IQon, Philips) and an SPCCT (Philips). The study was approved by the local ethical committee and patients signed an informed consent.
Objectives: To evaluate the image quality (IQ) of a spectral photon-counting CT (SPCCT) using filtered back projection (FBP) and hybrid iterative reconstruction (IR) algorithms (iDose), in comparison with a dual-layer CT (DLCT) system, and to choose the best image quality according to the IR level for SPCCT.
Methods: Two phantoms were scanned using a standard lung protocol (120 kVp, 40 mAs) with SPCCT and DLCT systems. Raw data were reconstructed using FBP and 9 iDose levels (i1/i2/i3/i4/i5/i6/i7/i9/i11) for SPCCT and 7 for DLCT (i1/i2/i3/i4/i5/i6/i7).
Purpose: The purpose of this study was to characterize the technical capabilities and feasibility of a large field-of-view clinical spectral photon-counting computed tomography (SPCCT) prototype for high-resolution (HR) lung imaging.
Materials And Methods: Measurement of modulation transfer function (MTF) and acquisition of a line pairs phantom were performed. An anthropomorphic lung nodule phantom was scanned with standard (120kVp, 62mAs), low (120kVp, 11mAs), and ultra-low (80kVp, 3mAs) radiation doses.
Objectives: When iodinated contrast is administered during CT simulation, standard practice requires a separate non-contrast CT for dose calculation. The objective of this study is to validate our hypothesis that since iodine affects Hounsfield units (HUs) more than electron density (ED), the information from post-contrast dual-layer CT (DLCT) would be sufficient for accurate dose calculation for both photon and proton therapy.
Methods And Materials: 10 pediatric patients with abdominal tumors underwent DLCT scans before and after iodinated contrast administration for radiotherapy planning.
Purpose: To assess whether virtual non-contrast (VNC) images derived from contrast dual-layer dual-energy computed tomography (DL-DECT) images could replace true non-contrast (TNC) images for aortic intramural hematoma (IMH) diagnosis in acute aortic syndrome (AAS) imaging protocols by performing quantitative as well as qualitative phantom and clinical studies.
Materials And Methods: Patients with confirmed IMH were included retrospectively in two centers. For in vitro imaging, a custom-made phantom of IMH was placed in a semi-anthropomorphic thorax phantom (QRM GmbH) and imaged on a DL-DECT at 120 kVp under various conditions of patient size, radiation exposure, and reconstruction modes.
Diagnostic imaging of hepatocellular carcinoma (HCC) requires a liver CT or MRI multiphase acquisition protocol. Patients would benefit from a high-resolution imaging method capable of performing multi-phase imaging in a single acquisition without an increase in radiation dose. Spectral Photon-Counting Computed Tomography (SPCCT) has recently emerged as a novel and promising imaging modality in the field of diagnostic radiology.
View Article and Find Full Text PDFBackground: To evaluate the feasibility of multicolour quantitative imaging with spectral photon-counting computed tomography (SPCCT) of different mixed contrast agents.
Methods: Phantoms containing eleven tubes with mixtures of varying proportions of two contrast agents (i.e.
Purpose: This study aimed to quantitate the accuracy of the determination of electron density (ED), effective atomic number (Z ), and iodine concentration, directed for more accurate radiation therapy planning, with a new dual-layer dual-energy computed tomography (DL-DECT) system. The dependence of the accuracy of these values on the scan and reconstruction parameters, as well as on the phantom size, was also examined.
Methods: Measurements were performed on a commercial DECT system with a DL detector (IQon Spectral CT, Philips Healthcare), using phantoms with various tissue-equivalent inserts as well as iodine and calcium inserts of different concentrations.
Purpose: Metal-oxide-semiconductor field-effect transistors (MOSFETs) serve as a helpful tool for organ radiation dosimetry and their use has grown in computed tomography (CT). While different approaches have been used for MOSFET calibration, those using the commonly available 100 mm pencil ionization chamber have not incorporated measurements performed throughout its length, and moreover, no previous work has rigorously evaluated the multiple sources of error involved in MOSFET calibration. In this paper, we propose a new MOSFET calibration approach to translate MOSFET voltage measurements into absorbed dose from CT, based on serial measurements performed throughout the length of a 100-mm ionization chamber, and perform an analysis of the errors of MOSFET voltage measurements and four sources of error in calibration.
View Article and Find Full Text PDFObjective: To estimate the potential dose reduction in iodinated contrast when interpreting monoenergetic images from spectral CT.
Methods: 51 paediatric patients received contrast-enhanced CT simulation for radiation therapy using a single-source, dual-layer detector spectral CT. The contrast-to-noise ratios (CNRs) of blood vessels were measured relative to surrounding soft tissue.