Publications by authors named "Yasuo Harata"

The objective of this study was to study the dose reduction effect on occupational exposure of operators and public exposure other than operators when a backscatter shield and a rectangular collimator are used in conjunction with a handheld intraoral X-ray unit. The occupational exposure was reduced to 40% when the backscatter shield was attached to the cone-tip, to 13% when the rectangular collimator was attached and to 7.7% when the backscatter shield and rectangular collimator were used together.

View Article and Find Full Text PDF

Portable intraoral X-ray units are frequently used for home-visit dental treatment and personal dental identification. Therefore, the reduction of operator exposure is crucial. Rectangular collimation is effective at reducing patient exposure and operator exposure; however, its effects are not known.

View Article and Find Full Text PDF

Objective: To investigate the dosage and imaging conditions for patients undergoing intraoral radiography at Meikai University Hospital and establish assurance and quality control data.

Methods: Tube voltage, exposure time, and air kinetic energy released per unit mass (air kerma) of three intraoral radiography units were measured. To calculate the patient entrance dose (PED) for each radiograph using Insight film, we extracted data for 1063 patients from their exposure records.

View Article and Find Full Text PDF