Objective: To explore the differences between oscillometric and auscultatory measurements.
Method: From a simulator evaluation of a non-invasive blood pressure (NIBP) device regenerating 242 oscillometric blood pressure waveforms from 124 subjects, 10 waveforms were selected based on the differences between the NIBP (oscillometric) and auscultatory pressure measurements. Two waveforms were selected for each of five criteria: systolic over and underestimation; diastolic over and underestimation; and close agreement for both systolic and diastolic pressures.
Oscillometric noninvasive blood pressure devices measure blood pressure using an indirect method and proprietary algorithms and hence require validation in clinical trials. Clinical trials are, however, expensive and give contradictory results, and validated devices are not accurate in all patient groups. Simulators that regenerate oscillometric waveforms promise an alternative to clinical trials provided they include sufficient physiological and pathological oscillometric waveforms.