A home-made chemical ionization time-of-flight mass spectrometer (TOFMS) has been developed for continuous measurements of atmospheric hydroxyl radical. Based on the atmospheric pressure chemical ionization technique, an ionization source with orthogonal dual tube structure was adopted in the instrument, which minimized the interference between the reagent gas ionization and the titration reaction. A 63Ni radioactive source was fixed inside one of the orthogonal tubes to generate reactant ion of NO(-)(3) from HNO3 vapor.
View Article and Find Full Text PDFWith the features of a broad range of ionizable compounds, reduced fragments and simple mass spectrum, a homemade magnetic field enhanced photoelectron ionization (MEPEI) source combined with single photon ionization (SPI) for time-of-flight mass spectrometer was built and applied to analyze thermal decomposition/combustion products of polyvinyl chloride (PVC). The combined ion source can be switched very fast between SPI mode and SPI-MEPEI mode for detecting different targeted compounds, and only adjusting the voltage of the electrode in the ionization region to trigger the switch. Among the PVC thermal decomposition/combustion products, HCl and CO2, which ionization energies (12.
View Article and Find Full Text PDFA home-made membrane inlet- single photon ionization/chemical ionization- time-of-flight mass spectrometer has been described. A vacuum ultraviolet (VUV) lamp with photon energy of 10.6 eV was used as the light source for single photon ionization (SPI).