Size exclusion (SE) high performance liquid chromatography (HPLC) is widely used for the molecular size distribution (MSD) analyses of various therapeutic proteins. We report development and validation of a SE-HPLC method for MSD analyses of immunoglobulin G (IgG) in products using a TSKgel SuperSW3000 column and eluting it with 0.4M NaClO, a chaotropic salt, in 40mM phosphate buffer, pH 6.
View Article and Find Full Text PDFWe report development and validation of a simple, rapid, and accurate method for the quantitation of protein nitrogen, which combines Kjeldahl digestion and ion chromatography with suppressed conductivity detection and requires nanomolar amount of nitrogen in samples (≥10 μg protein). The mechanism of suppressed conductivity detection does not permit analysis of samples containing copper (present in Kjeldahl digestion solution) and aluminum (present in many vaccines as adjuvants) due to precipitation of their hydroxides within the suppressor. We overcame this problem by including 10 μM oxalic acid in Kjeldahl digests and in the eluent (30 mM methanesulfonic acid).
View Article and Find Full Text PDFMost biological products, including vaccines, administered by the parenteral route are required to be tested for sterility at the final container and also at various stages during manufacture. The sterility testing method described in the Code of Federal Regulations (21 CFR 610.12) and the United States Pharmacopoeia (USP, Chapter <71>) is based on the observation of turbidity in liquid culture media due to growth of potential contaminants.
View Article and Find Full Text PDF