Background: In the second generation of the point-of-care (POC) assay Roche CARDIAC proBNP, the upper limit of the measuring range was extended from 3000 to 9000 ng/L.
Methods: A thirteen-site multicentre evaluation was carried out to assess the analytical performance of the POC NT-proBNP assay and to compare it with a laboratory N-terminal pro-brain natriuretic peptide (NT-proBNP) assay.
Results: In method comparisons of six lots of POC NT-proBNP with the lab reference method (Elecsys proBNP) mean bias ranged from -10 to +17%.
A new point-of-care test for the determination of NT-proBNP in whole blood was developed based on the existing gold-label rapid immunoassay technology of the Roche Cardiac reader system. The novel gold-labelled monoclonal antibody recognizes NT-proBNP at amino acid sequence 27 to 31, the biotinylated polyclonal antibody recognizes sequence 39 to 50. In a model assay based upon the reference method Elecsys proBNP and with an R & D lot of the point-of-care test, this newly selected and developed combination of antibodies showed a very good correlation with the standard Elecsys proBNP assay with correlations of 0.
View Article and Find Full Text PDFBackground: The Roche CARDIAC proBNP point-of-care (POC) test is the first test intended for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in whole blood as an aid in the diagnosis of suspected congestive heart failure, in the monitoring of patients with compensated left-ventricular dysfunction and in the risk stratification of patients with acute coronary syndromes.
Methods: A multicentre evaluation was carried out to assess the analytical performance of the POC NT-proBNP test at seven different sites.
Results: The majority of all coefficients of variation (CVs) obtained for within-series imprecision using native blood samples was below 10% for both 52 samples measured ten times and for 674 samples measured in duplicate.