Estimating radiation doses for operators performing interventional radiological procedures is crucial in the occupational radiation protection of medical staff. In this study, Monte Carlo simulations coupled with an anthropomorphic phantom were used to model various exposure scenarios during the procedures. Conversion coefficients of the dose-area product of x rays for the eye lens equivalent dose, hand equivalent dose, and whole-body effective dose of the operator were calculated.
View Article and Find Full Text PDFCd is quite a useful nuclide for the calibration of X-ray spectroscopy, gamma-ray spectroscopy or as an excitation source in X-ray fluorescence. In this work, 4π(PC)e-X coincidence counting and 4π(LS)ce counting were used to determine the activity of the Cd solution. The two methods showed good agreement in their counting results.
View Article and Find Full Text PDFThe National Radiation Standard Laboratory of Taiwan organized in 2014 a comparison exercise by distributing 210 L drum-typed samples to seven radioactive waste analysis laboratories in Taiwan. Four drums were filled with uniformly distributed active carbon, water, resin and concrete, respectively and five drums were filled with cracked metals and heterogeneously distributed radioactive sources. Measurement uncertainties of participants results are in the range 3–40% (k=2) and about 96% of the reported results produced En values (ISO, 1997) smaller than one for drums with activity uniformly distributed.
View Article and Find Full Text PDF