Publications by authors named "Walter Russ"

Introduction: Immediate accessibility to automated external defibrillators (AED) is recommended for highly frequented public areas. In train terminals and metro stations electromagnetic interference (EMI) is present. In preparation for a public access defibrillation (PAD) programme in this environment possible effects on AED safety and accuracy were studied.

View Article and Find Full Text PDF

Objectives: Emergency services personnel, family members, laypersons or patients often carry and use mobile phones on sites of emergencies. As there are reported effects on implanted pacemakers and cardioverter defibrillators, the influence of digital cellular phones on automated external defibrillators was studied.

Methods: Twelve automated external defibrillator models were bench tested for their correct decision to or not to advise a shock, while being exposed to electromagnetic interference from a handheld cellular phone with 2 W or a portable cellular phone with 8 W transmitting power.

View Article and Find Full Text PDF