Conventionally, a dual-end depth-of-interaction (DOI) block detector readout requires two two-dimensional silicon photomultiplier (SiPM) arrays, one on top and one on the bottom, to define the XYZ positions. However, because both the top and bottom SiPM arrays are reading the same pixels, this creates information redundancy. We propose a dichotomous orthogonal symmetric (DOS) dual-end readout block detector design, which removes this redundancy by reducing the number of SiPMs and still achieves XY and DOI (Z) decoding for positron emission tomography (PET) block detector.
View Article and Find Full Text PDFAccurate PET system timing alignment minimizes the coincidence time window and therefore reduces random events and improves image quality. It is also critical for time-of-flight (TOF) image reconstruction. Here, we use a thin annular cylinder (shell) phantom filled with a radioactive source and located axially and centrally in a PET camera for the timing alignment of a TOF PET system.
View Article and Find Full Text PDFUnlabelled: The dedicated murine PET (MuPET) scanner is a high-resolution, high-sensitivity, and low-cost preclinical PET camera designed and manufactured at our laboratory. In this article, we report its performance according to the NU 4-2008 standards of the National Electrical Manufacturers Association (NEMA). We also report the results of additional phantom and mouse studies.
View Article and Find Full Text PDFA fully digital FPGA-based high count-rate coincidence system has been developed for TOF (Time of Flight) and non-TOF PET cameras. Using a hybrid of AND-logic and Time-mark technology produced both excellent timing resolution and high processing speed. In this hybrid architecture, every gamma event was synchronized by a 125 MHz system clock and generating a trigger associated with a time-mark given by an 8-bit high-resolution TDC (68.
View Article and Find Full Text PDFA lower-cost high-sensitivity high-resolution positron emission mammography (PEM) camera is developed. It consists of two detector modules with the planar detector bank of 20 × 12 cm(2). Each bank has 60 low-cost PMT-Quadrant-Sharing (PQS) LYSO blocks arranged in a 10 × 6 array with two types of geometries.
View Article and Find Full Text PDFThe goal of this work is to develop a novel, accurate, real-time digital baseline restorer using online statistical processing for a high count-rate digital system such as positron emission tomography (PET). In high count-rate nuclear instrumentation applications, analog signals are DC-coupled for better performance. However, the detectors, pre-amplifiers and other front-end electronics would cause a signal baseline drift in a DC-coupling system, which will degrade the performance of energy resolution and positioning accuracy.
View Article and Find Full Text PDFIn photomultiplier-quadrant-sharing (PQS) geometry for positron emission tomography applications, each PMT is shared by four blocks and each detector block is optically coupled to four round PMTs. Although this design reduces the cost of high-resolution PET systems, when the camera consists of detector panels that are made up of square blocks, half of the PMT's sensitive window remains unused at the detector panel edge. Our goal was to develop a LYSO detector panel which minimizes the unused portion of the PMTs for a low-cost, high-resolution, and high-sensitivity positron emission mammography (PEM) camera.
View Article and Find Full Text PDFWe developed a detailed Monte Carlo simulation method to study the time resolution of detectors for time-of-flight positron emission tomography (TOF PET). The process of gamma ray interaction in detectors, scintillation light emission and transport inside the detectors, the photoelectron generation and anode signal generation in the photomultiplier tube (PMT), and the electronics process of discriminator are simulated. We tested this simulation method using published experimental data, and found that it can generate reliable results.
View Article and Find Full Text PDFWe developed high resolution L(Y)SO detectors for human and animal PET applications using Photomultiplier-quadrant-sharing (PQS) technology. The crystal sizes were 1.27 × 1.
View Article and Find Full Text PDF